首页> 外文期刊>Experimental Mechanics >Characterization of a Laser Extensometer for Split Hopkinson Pressure bar Experiments
【24h】

Characterization of a Laser Extensometer for Split Hopkinson Pressure bar Experiments

机译:用于分裂霍普金森压杆实验的激光突出仪的特征

获取原文
获取原文并翻译 | 示例
           

摘要

This paper examines a laser extensometer to calculate the strain in a sample during split Hopkinson pressure bar (SHPB) experiments. This method offers a non-contact, direct method for measuring sample strain which does not rely on one-dimensional wave propagation assumptions. First a single bar experiment is used to compare the extensometer's accuracy and frequency response against a laser vibrometer and an accelerometer. The extensometer showed a close match with the vibrometer up to a bandwidth of 10 kHz. With the performance validated, the extensometer is applied to a SHPB experiment with silicone samples. For low strains, the extensometer shows a good match to the strain determined by strain gauges on the bars. At higher strains, the radial expansion of the sample can interfere with the measurement beam from the extensometer and produce inaccurate results.
机译:本文介绍了激光突出仪,以在分裂霍普金森压力棒(SHPB)实验期间计算样品中的应变。 该方法提供了用于测量样品应变的非接触式直接方法,其不依赖于一维波动传播假设。 首先,单个条形实验用于比较伸展仪的精度和频率响应对激光振动器和加速度计。 伸展计显示与振动计的紧密匹配,直到10 kHz的带宽。 通过验证性能,扩展仪应用于具有硅胶样品的SHPB实验。 对于低菌株,伸展计显示与由杆上的应变计确定的应变匹配的良好匹配。 在较高的菌株处,样品的径向膨胀可以干扰来自突出计的测量光束并产生不准确的结果。