Background: During intensity modulated radiation therapy (IMRT) technique, theoretically, presence of flattening filter (FF) across the beamline of clinical linear accelerator (linac) is not essential to obtain uniform lateral profiles due to intensity modulation of photon beams by multileaf collimators (MLCs). The aim of this study was to investigate the dosimetrical properties of 18 MV photon beam-Varian linac with and without FF. Materials and Methods: All dose measurements were performed on 18 MV, FF mode-Varian 2100C/D linac. The FF and flattening filter free (FFF) modes of linac were modeled by MCNPX 2.4. code. The photon and contaminant electrons spectra, dose rate, present depth doses (PDD), lateral dose profiles, total and collimator scatter factors and out of field doses were calculated and compared with and without FF. Results: Removing the FF increased the photon and contaminant electron fluences by factors of 5.48 and 3.94 for a 5 x 5 cm2 field size, respectively. The surface dose increased up to 155%. The flatness of dose profile was disturbed and deteriorated with increase of field size. Despite the dependence of the total scattering factor on field size, the variation of collimator scattering factors was neglected. The out-of-field dose decreased about 81.5 % for a 5 x 5 cm2 field size. Conclusion: Removing FF from the linac head increases the dose rate and decreases the out-of-field dose, but the increased skin dose and deteriorated flatness of lateral dose profile are the main disadvantages of the FFF mode.
机译:背景:在强度调制的放射治疗(IMRT)技术期间,理论上,在临床线性加速器(LINAC)的光束线上存在扁平滤波器(FF)的存在不是必不可少的,以获得由于Multireaf准直器的光子束的强度调制而获得均匀的横向型材( MLCS)。本研究的目的是研究18 mV光子束 - 瓦拉的微量特性,有没有FF。材料和方法:在18 mV,FF模式 - Varian 2100C / D LINAC上进行所有剂量测量。 LINAC的FF和扁平化滤波器(FFF)模式由MCNPX 2.4建模。代码。计算光子和污染电子光谱,剂量率,本深度剂量(PDD),横向剂量分析,总和准直器散射因子以及与野外剂量超出阳离子,与并且没有FF。结果:通过5.48和3.94的因子,除去FF增加光子和污染电子流量,分别为5×5cm2场大小。表面剂量增加高达155%。随着场尺寸的增加,剂量曲线的平坦度受到干扰并劣化。尽管总散射因子对场尺寸的依赖性,但忽略了准直器散射因子的变化。场外剂量为5×5cm 2场大小降低约81.5%。结论:从LinaC头中除去FF增加剂量率并降低外场剂量,但增加的皮肤剂量和横向剂量曲线的劣化是FFF模式的主要缺点。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号