A comparison of contralateral breast dose due to breast cancer radiotherapy using two different treatment machines in a radiotherapy center

机译:放射治疗中心两种不同处理机带来乳腺癌放射治疗引起的对侧乳腺剂量的比较

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Background: The radiation dose received by contralateral breast (CLB) is one of the concerns of breast radiotherapy, because it may lead to the induction of secondary breast cancer. The aim of this study was to evaluate the CLB surface dose in the breast treatment in Yazd radiotherapy center. Materials and Methods: The surface dose of CLB was measured using TLD dosimetry in 50 cancer breast patients. The TLD chips were placed at four points on the each of CLBs. The patients were treated by 6MV photon beams of Oncor (physical wedge) and Compact (motorized wedge) LINAC. The TLD chips were placed on the surfaces of CLB during the medial and lateral tangent radiation fields in one of radiotherapy fractions. Results: The mean percent of prescription dose of the CLB surface doses on the point 1 in the two Linac (Oncor & Compact) were significantly different. The mean of CLB surface doses of point 1 in the physical and the motorized wedge techniques were 5.78 and 7.84 percent of prescription dose of breast cancer, respectively. The medial and lateral fields' contribution from 7.4% surface dose of CLB were 5.8% and 1.6%, respectively. Conclusion: In Shahid Ramezanzadeh radiotherapy center, the CLB surface dose due to breast cancer radiotherapy by the Compact machine (7.84 %) was significantly more than the allowable value (6% prescription dose). The CLB does due to the medial field beam was more than the lateral field.
机译:背景:对侧乳房(CLB)接收的辐射剂量是乳房放射治疗的关注之一,因为它可能导致诱导次级乳腺癌。本研究的目的是评估亚兹德放射治疗中心乳房治疗中的CLB表面剂量。材料和方法:使用50例癌症乳腺患者的TLD剂量测定来测量CLB的表面剂量。将TLD芯片放在CLB中的每一个上的四个点。患者用6MV光子梁的oncor(物理楔形)和紧凑型(机动楔形)LINAC处理。将TLD芯片放置在放射治疗部分之一中的内侧和横向切线辐射区域的CLB的表面上。结果:两个LINAC(oncor&Compact)的点1中CLB表面剂量的平均百分比百分比显着不同。物理和电动楔技术中CLB表面剂量的CLB表面剂量的平均值分别为5.78%和7.84%的处方剂量的乳腺癌。从7.4%的CLB的内侧和横向场的贡献分别为5.8%和1.6%。结论:在Shahid Ramezanzadeh放射疗法中心,Compact机器(7.84%)引起的CLB表面剂量由于乳腺癌放射治疗(7.84%)比允许值(6%处方剂量)显着大。由于内侧场梁的CLB是多于横向场。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号