Purpose To evaluate the results of optical biometry using the IOLMaster 500, Lenstar LS 900 and Aladdin in eyes with cataract. Methods In 231 eyes of 152 patients with cataract, the measurements of 3 different biometers were retrospectively compared. Paired comparisons were performed for axial length (AL), mean keratometry (mean K) and anterior chamber depth (ACD). Results In only 197 of the 231 eyes (85.3%), it was possible to obtain reliable measurements of AL with all the three devices. It was not possible to determine AL in 16 eyes (6.9%) with Lenstar LS 900; in 19 eyes (8.2%) with Aladdin; and in 20 eyes (8.6%) with IOLMaster 500 possibly related to the severity of lens opacification (the corneas had good transparency in the eyes included in the study). There was a statistically significant difference in AL between IOLMaster 500 and the remaining two biometers (P = 0.03). However, the amount of difference was considered clinically not significant (0.04 mm). The mean ker-atometry (mean K) was determined in 203 eyes (87.9%) with all the three devices. Differences in mean K were between — 0.1 and 0.06 Diopters (D), which were considered neither statistically (P > 0.05) nor clinically significant. The anterior chamber depth (ACD) was determined in 197 eyes (85.28%) with all the three biometers. The differences between the three devices (0.03 to 0.13 mm) were not statistically significant and considered also clinically not significant. Conclusions There were no clinically significant differences between these 3 biometers in AL, mean K and ACD.
机译:目的使用Iolmaster 500,Lenstar LS 900和Aladdin与白内障的眼睛中的光学生物测量结果评估结果。方法231只白内障患者的231只眼睛,测量3种不同的生物座位的测量比较。对轴向长度(Al)进行配对的比较,平均变音测定(平均值K)和前房深度(ACD)。结果仅在231只眼睛(85.3%)中,可以获得所有三种器件的可靠测量。通过Lenstar LS 900,不可能在16只眼睛(6.9%)中确定Al; 19只眼睛(8.2%)与阿拉丁;在20只眼里(8.6%)与Iolmaster 500可能与镜片渗透率的严重程度有关(角膜在研究中的眼中有良好的透明度)。在IOLMASTER 500和剩余的两个生物座位之间存在统计学上的显着差异(P = 0.03)。然而,差异的差异被认为是临床上不显着的(0.04mm)。用所有三个装置在203只眼睛(87.9%)中测定平均Ker-测量法(平均值K)。平均值的差异介于0.1和0.06屈光度(D)之间,其既不统计(P> 0.05)也不临床显着。前房深度(ACD)在197只眼睛(85.28%)中确定了所有三个生物夹。三种器件(0.03至0.13mm)之间的差异在统计学上没有统计学意义,并且在临床上被认为是不显着的。结论在Al中,这3个生物夹之间没有临床上显着差异,平均值K和ACD。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号