Reliability and validity of two software systems used to measure the pharyngeal airway space in three-dimensional analysis

机译:用于测量三维分析中咽部气道空间的两个软件系统的可靠性和有效性

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The aim of this study was to test the reliability and validity of two software systems used to measure the pharyngeal airway space three-dimensionally. A sample of 40 cone beam computed tomography images from adult patients was taken from a database. The cone beam computed tomography images were analysed by InVivoDental and Dolphin 3D software systems by two calibrated examiners. Three nasopharynx and oropharynx prototypes were used as a reference standard to validate the software systems. The volume, minimum area and minimum area localization were the measurements tested. Measurements were compared using a paired t-test; correlated using Pearson's correlation and linear regression. Bland- Altman analysis was also used. We found significant differences in the oropharynx volume (P = 0.002) and nasopharynx minimum area localization (P = 0.009). The Dolphin 3D software presented higher-volume values than the ones found in the prototype, while the InVivoDental software presented lower values. Strong (r>0.7; P > 0.001) or very strong (r>0.9; P > 0.001) correlations were observed between the software systems. Bland-Altman analysis found good agreement between prototypes and the software systems. The measurements obtained from the Dolphin 3D and InVivoDental software systems are both reliable, strongly correlated, but should not be assumed as equal. Dolphin 3D software overestimates the nasopharynx and oropharynx volumes, while the InVivoDental software underestimates them.
机译:本研究的目的是测试用于三维测量咽部气道空间的两个软件系统的可靠性和有效性。来自成年患者的40个锥形光束计算机断层扫描图像的样本从数据库中获取。通过两次校准检查员通过辅助和海豚3D软件系统分析了锥形光束计算机断层摄影图像。使用三个鼻咽和oropharynx原型作为参考标准来验证软件系统。体积,最小面积和最小区域定位是测试的测量值。使用配对的T检验进行测量;使用Pearson的相关性和线性回归相关。也使用了Bland-Altman分析。我们发现Oropharynx体积的显着差异(p = 0.002)和鼻咽最小区域定位(p = 0.009)。海豚3D软件呈现比原型中的更高卷的值,而邀请软件呈现较低的值。在软件系统之间观察到强(r> 0.7; p> 0.001)或非常强(r> 0.9; p> 0.001)相关性。 Bland-Altman分析发现原型与软件系统之间的良好一致。从Dolphin 3D和vivovental软件系统获得的测量既可靠,强烈相关,但不应被认为是相等的。海豚3D软件高估了鼻咽和oropharynx卷,而invivodents软件低估了它们。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号