首页> 外文期刊>International journal of laboratory hematology >Stability of complete blood count in different storage conditions using the ABX PENTRA ABX PENTRA 60 analyzer
【24h】

Stability of complete blood count in different storage conditions using the ABX PENTRA ABX PENTRA 60 analyzer

机译:使用ABX PENTRA ABX PETRA 60分析仪在不同储存条件下完全血液计数的稳定性

获取原文
获取原文并翻译 | 示例
       

摘要

Abstract Introduction Sample stability is essential to obtain reliable results in the clinical laboratory. This study was conducted to investigate the reliability of hematological parameters using ABX Pentra 60 in samples stored for up to 72?hours at different temperatures. Methods A total of 651 blood samples were analyzed at different analysis times: 2, 24, 48, and 72?hours and forms and storage: room temperature (25°C) and at 4°C. The imprecision of the results was evaluated by the analytical coefficient of variation ( CV a%) obtained by the typical error ( TE ) and Kruskal‐Wallis analysis, to compare the averages. The reliability of the results was evaluated by the CV a (%) within the maximum allowable analytical variation and by the difference of means of the results in relation to the baseline sample (2?hours). Results Red blood count, hemoglobin, and MCH parameters showed stability up to 72?hours at room temperature and at 4°C. The other complete blood count parameters showed imprecision results emitted by the ABX Pentra 60 from 24?hours of sample storage, independent of the storage temperature. In addition, there were significant oscillations in the mean values, particularly for the samples stored at room temperature, with the exception of platelet parameters that exhibited mean changes also at 4°C. Conclusion The results demonstrate the importance of the clinical analyst's knowledge about the behavior of the CBC parameters over time under different storage conditions, and mainly the imprecision of the hematological equipment used, for the suitable interpretation of the results.
机译:摘要引言样本稳定性对于获得临床实验室的可靠结果至关重要。进行该研究以研究使用ABX PENTRA 60在不同温度下储存最高72小时的样品中的血液学参数的可靠性。方法在不同的分析时间分析总共651次血样:2,24,48和72℃,形成和储存:室温(25℃)和4℃。通过典型误差(TE)和Kruskal-Wallis分析获得的分析系数(CV A%)来评估结果的不精确评估,以比较平均值。结果的可靠性由CV A(%)在最大允许的分析变化中和通过与基线样品(2Ω小时)相关的方法的差异来评估。结果红血计数,血红蛋白和MCH参数在室温下稳定至72小时,4°C。另一个完整的血统计数参数显示ABX PENTRA 60从24小时内发出的不确定结果,与储存温度无关。另外,平均值中存在显着的振荡,特别是对于在室温下储存的样品,除了在4℃下表现出平均变化的血小板参数外。结论结果表明,临床分析师在不同储存条件下随着时间的推移,临床分析师对CBC参数的行为的重要性,主要是使用血液学设备的不精确,用于对结果的合适解释。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号