The present study aimed to investigate the comparability of capillary electrophoresis (CE) and three different methods for HbA1c measurement. The 270 whole blood samples with normal hemoglobin profiles were analyzed for HbA1c values by the Roche turbidimetric inhibition immunoassay (TINIA), the Mindray enzymatic assay (EA), the Arkray high-performance liquid chromatography (HPLC) in comparison with the Sebia CE. The within-laboratory coefficient of variation of all four methods was within the accepted goal (<2%), showing good performance of all of these methods. Pairwise comparisons of HbA1c values obtained by CE and other methods were determined in both total group and subgroups (HbA1c levels of <6.5%, 6.5-8% and >8%). Mean differences of HbA1c values in all groups were very small in which the mean HbA1c value measured by EA was lower while those by TINIA and HPLC were higher than that of CE. The majority of different values were within the limits of agreement in Bland-Altman analysis, indicating good agreement between CE and the others. Less than 5% of percentage differences were out of the total allowable error limit in all groups, showing that differences of HbA1c values between CE and other methods were not clinically significant. Pairwise comparisons of HbA1c values of CE and the others in Passing-Bablok regression and Spearman rank correlation studies displayed high concordance and strong correlation in all groups. In conclusion, the present study showed strong correlation, high comparability and consistent results for HbA1c measurement between capillary electrophoresis and the other three different methods.
机译:本研究旨在研究毛细管电泳(CE)和三种不同方法对HBA1C测量的可比性。通过罗氏浊度抑制免疫测定(Tinia)分析了具有正常血红蛋白型材的270个具有正常血红蛋白曲线的血液样品,与Sebia Ce相比,MindRay酶法测定(EA),Arkray高性能液相色谱(HPLC)。所有四种方法的实验室内变异系数在接受的目标(<2%)内,显示出所有这些方法的良好性能。通过Ce和其他方法获得的HBA1C值的成对比较在总基团和亚组(HBA1C水平<6.5%,6.5-8%和> 8%)中测定。所有基团中HBA1C值的平均值非常小,其中EA测量的平均HBA1C值较低,而细胞和HPLC的均是高于CE的那些。大多数不同的价值观在Bland-Altman分析中的协议范围内,表明CE与其他人之间的良好一致。百分比差异的百分比差异不到所有组中的总允许误差限制,表明CE和其他方法之间的HBA1C值差异在临床上没有临床显着性。 CE和Spearman等级相关研究中CE和其他其他人的HBA1C值的成对比较显示了所有群体的高一致性和强烈的相关性。总之,本研究表明毛细管电泳与其他三种不同方法之间的HBA1C测量的相关性,高可比性和一致结果。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号