首页> 外文期刊>Acta Radiologica >Automatic exposure control in computed tomography--an evaluation of systems from different manufacturers.
【24h】

Automatic exposure control in computed tomography--an evaluation of systems from different manufacturers.

机译:计算机断层摄影中的自动曝光控制-对来自不同制造商的系统的评估。

获取原文
获取原文并翻译 | 示例
       

摘要

BACKGROUND: Today, practically all computed tomography (CT) systems are delivered with automatic exposure control (AEC) systems operating with tube current modulation in three dimensions. Each of these systems has different specifications and operates somewhat differently. PURPOSE: To evaluate AEC systems from four different CT scanner manufacturers: General Electric (GE), Philips, Siemens, and Toshiba, considering their potential for reducing radiation exposure to the patient while maintaining adequate image quality. MATERIAL AND METHODS: The dynamics (adaptation along the longitudinal axis) of tube current modulation of each AEC system were investigated by scanning an anthropomorphic chest phantom using both 16- and 64-slice CT scanners from each manufacturer with the AEC systems activated and inactivated. The radiation dose was estimated using the parameters in the DICOM image information and image quality was evaluated based on image noise (standard deviation of CT numbers) calculated in 0.5 cm(2) circular regions of interest situated throughout the spine region of the chest phantom. RESULTS: We found that tube current modulation dynamics were similar among the different AEC systems, especially between GE and Toshiba systems and between Philips and Siemens systems. Furthermore, the magnitude of the reduction in the exposure dose was considerable, in the range of 35-60%. However, in general the image noise increased when the AEC systems were used, especially in regions where the tube current was greatly decreased, such as the lung region. However, the variation in image noise among images obtained along the scanning direction was lower when using the AEC systems compared with fixed mAs. CONCLUSION: The AEC systems available in modern CT scanners can contribute to a significant reduction in radiation exposure to the patient and the image noise becomes more uniform within any given scan.
机译:背景技术:今天,几乎所有计算机断层扫描(CT)系统都带有自动曝光控制(AEC)系统,该系统通过三维三维电流调制工作。这些系统中的每一个都有不同的规格,并且操作有所不同。目的:评估来自四个不同的CT扫描仪制造商的AEC系统:通用电气(GE),飞利浦,西门子和东芝,考虑它们在减少辐射量的同时保持适当图像质量的潜力。材料与方法:通过激活和停用AEC系统的每个制造商的16层和64层CT扫描仪,通过扫描拟人化的人体模型来研究每个AEC系统的电流调制的动力学(沿纵轴的适应性)。使用DICOM图像信息中的参数估计辐射剂量,并根据在整个胸部幻影的脊柱区域中放置的0.5 cm(2)圆形感兴趣区域中计算出的图像噪声(CT数的标准差)评估图像质量。结果:我们发现,在不同的AEC系统之间,特别是在GE和东芝系统之间以及在Philips和Siemens系统之间,管电流调制动力学是相似的。此外,暴露剂量减少的幅度是可观的,在35-60%的范围内。但是,通常,当使用AEC系统时,图像噪声会增加,特别是在显像管电流大大降低的区域(例如肺区域)中。但是,使用AEC系统时,与固定mAs相比,沿扫描方向获得的图像之间的图像噪声变化较小。结论:现代CT扫描仪中可用的AEC系统可以显着减少对患者的辐射暴露,并且在任何给定的扫描范围内图像噪声变得更加均匀。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号