Comparative study of the serum measurement of PTH on Roche Cobas e411 (R) versus the Abbott Architect ci8200 (R)

机译:罗氏罗氏玉米棒型欧洲胸骨血清测量比较研究与Abbott Architect Ci8200(R)对比

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Parathormone (PTH) is the main hormone of phosphocalcic homeostasis. It is synthesized and secreted by the parathyroid glands. PTH has become a routine test in the medical biology laboratory. However, its measurement presents analytical difficulties with the various marketed kits. The aim of this work is to present the results of a comparative study between the PTH measurment on Abbott architect ci8200 and on Roche's Cobas e411 automaton. It is a prospective study carried out for 252 hospitalized patients in the various departments of the University Hospital Center Mohammed VI of Oujda. The "intact" PTH tests were performed on two automata: Abbott Architect ci8200 and Roche Cobas e411. The first uses chemiluminescent microparticle immunoassay. The second uses electrochemiluminiscence sandwich enzyme immunoassay. The agreement of the results between the different techniques was evaluated using the Bland-Altman difference diagram and the Passing-Bablok and Deming regression line (MedCalc software version 14.8.1.0 (R)). By analyzing the diagram of Bland-Altman, we note that the average bias between both methods is of the order of 193.9 pg/mL. As for the equation of the right of Passing-Bablok, it is: Y(Architect) = 3.11 X (Cobas) - 12.26. In conclusion, our study shows a great discrepancy between the results of the PTH assay on the Architect ci8200 versus the Cobas e411, hence the biologist's indisputable role in the control and evaluation of the kits marketed through the various validation tests.
机译:屈曲态(PTH)是磷酸钙稳态的主要激素。它由甲状旁腺合成和分泌。 PTH已成为医学生物实验室的常规测试。然而,其测量与各种销售套件具有分析困难。这项工作的目的是展示Abbott Architect CI8200和Roche的COBAS E411自动机的PTH测量比较研究结果。它是在大学医院中心穆罕默德州穆罕默德州穆罕默德六世各部门的252名住院患者患者进行了一项前瞻性研究。 “完整”PTH测试是对两种自动机进行的:Abbott Architect Ci8200和Roche Cobas E411进行。第一次使用化学发光微粒免疫测定。第二种用途电化学inisciscence夹心酶免疫测定。使用Bland-Altman差异图和通过-Bablok和Deming回归线(Medcalc软件版本14.8.1.0(r))评估不同技术之间的结果的协议。通过分析Bland-Altman的图,我们注意到两种方法之间的平均偏压是193.9 pg / ml的顺序。至于通过BABLOK的权利等式,它是:Y(架构师)= 3.11 x(COBAS) - 12.26。总之,我们的研究表明,在建筑师CI8200与COBAS E411对COBAS E411的结果结果之间存在巨大差异,因此生物学家在通过各种验证测试中销售的套件的控制和评估中的无可争议的作用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号