The number of patients with implanted left ventricular assist devices is constantly increasing. Numerous patients suffer from drug-resistant ventricular tachycardias. Only a few studies and reports about ablation therapy in this cohort of patients are available. Electromagnetic interferences between the left ventricular assist device and the three-dimensional mapping system which is used for ablation has been described as disabling the investigator to create a proper map of the left ventricle. We observed that this interference is interrupted by hot mapping meaning that one pretends an ablation with only 5 W, permitting the creation of a clear map in areas of interference.
机译:植入左心室辅助装置的患者的数量不断增加。 众多患者患有耐药性心室性心动过速。 只有少数关于这种患者队列的消融治疗的研究和报告。 左心室辅助装置和用于消融的三维映射系统之间的电磁干扰已被描述为禁用研究者创建左心室的适当地图。 我们观察到,这种干扰被热映射中断,这意味着一个只用5 W假装消融,允许在干扰区域中创建清晰的地图。