首页> 外文期刊>Archives of orthopaedic and trauma surgery. >Intraosseous correction of misdirected cannulated screws and fracture malalignment using a bent tip 2.0 mm guidewire: technique and indications
【24h】

Intraosseous correction of misdirected cannulated screws and fracture malalignment using a bent tip 2.0 mm guidewire: technique and indications

机译:使用弯曲尖端2.0 mm导丝骨膜矫正误导式插管和断裂污垢:技术和适应症

获取原文
获取原文并翻译 | 示例
           

摘要

Percutaneous pelvic screw placement is a technically demanding procedure. A precise intraosseous pathway must be prepared before screw placement into any osseous fixation pathway of the pelvis. Adjustments to a drill or guidewire become increasingly difficult as the instrument is advanced within the pelvis. We present a reliable and reproducible technique using a 2.0 mm guidewire that allows for correction of an initially misdirected drill within the pelvis. This technique also allows for manipulation and reduction of certain malaligned pelvic fractures prior to percutaneous cannulated screw placement. This technique does not substitute for poor surgical technique but is used to optimize the position of percutaneously placed pelvic screws.
机译:经皮骨盆螺杆放置是一种技术上要求苛刻的程序。 必须在将螺钉放置到骨盆的任何骨固定途径之前制备精确的骨内途径。 当仪器在骨盆内先进时,对钻头或导丝的调整变得越来越困难。 我们使用2.0mm导丝呈现可靠和可重复的技术,其允许校正骨盆内最初误导的钻头。 该技术还允许在经皮插管螺钉放置之前操纵和减小某些矿物骨盆裂缝。 该技术不替代手术技术差,但用于优化Percutane地放置盆腔螺钉的位置。