Effective and High-Throughput Analytical Methodology for the Determination of Lead and Cadmium in Water Samples by Disposable Pipette Extraction Coupled with High-Resolution Continuum Source Graphite Furnace Atomic Absorption Spectrometry (HR-CS GF AAS)

机译:通过一次性移液管萃取与高分辨率连续源石墨炉原子吸收光谱法(HR-CS GF AAS)相结合的有效和高通量分析方法。

获取原文
获取原文并翻译 | 示例
           

摘要

A new procedure for the preconcentration of trace amounts of free Pb and Cd by disposable pipette extraction (DPX) is proposed herein. Recycled cork is used as a biosorbent and the procedure is completely free of organic solvents. The cork was reduced to a powder and characterized by scanning electron microscopy. Several parameters that influence the preconcentration of Pb and Cd with DPX-cork, such as the sample pH, number of extraction cycles, biosorbent mass, and percentage of acid in the desorption step, were studied. The tolerance of DPX-cork with respect to 10 potential interfering ions was also evaluated. Coupled with the high-resolution continuum source graphite furnace atomic absorption spectrometry (HR-CS GF AAS) technique, the sample preparation method allowed the analytical limits required for quality control (according to the permitted limits established by legislation) to be obtained. The limits of detection for the extraction of a 3.5-mL water sample were 200 ng L-1 for Pb and 100 ng L-1 for Cd. The relative standard deviations were around 7.5% for Pb and 8.0% for Cd. The optimized method was successfully applied to the determination of Cd and Pb in two certified reference materials (water and wastewater) and five water samples (collected from a mangrove, a creek, and the sea).
机译:本文提出了一种通过一次性移液管萃取(DPX)的痕量游离Pb和Cd的前浓度的新方法。再循环软木用作生物吸附剂,程序完全没有有机溶剂。将软滤成软滤成粉末并通过扫描电子显微镜表征。研究了诸如解吸步骤中的诸如样品pH,萃取循环,生物吸道性循环,生物吸附循环,生物吸收性质量和酸百分比的若干参数。还评价了DPX-CORK相对于10个潜在干扰离子的容差。再加上高分辨率连续源石墨炉原子吸收光谱法(HR-CS GF AAS)技术,采样制备方法允许获得质量控制所需的分析限制(根据立法建立的允许限额)。用于3.5ml水样的检测限率为Pb和100ng L-1的200ng L-1。 Pb的相对标准偏差约为7.5%,CD为8.0%。优化的方法成功地应用于两种认证的参考资料(水和废水)和五种水样(从红树林,一条小溪和海洋收集)的CD和Pb的测定。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号