机译:紧凑型术中MRI系统进行颅骨手术和导航。

获取原文
获取原文并翻译 | 示例
           

摘要

Experience with a compact unit for intraoperative magnetic resonance imaging (iMRI) and integrated surgical navigation is presented. The system, commercially known as the PoleStar N-10 (Odin Medical Technologies, Yokne'am, Israel) includes a 0.12 Tesla permanent magnet that docks under a regular OR table. A passive infrared-based optical navigation system is included and the device can be controlled by the operating neurosurgeon and staff. We have operated on 93 patients using the PoleStar N-10. Diagnoses included glioma in 27, pituitary adenoma in 19, meningioma in 15, and others in 40. Lesions were removed from the skull base in 35 patients of whom 8 had posterior fossa masses. Surgery was affected by imaging in 51% of operations. Additional lesion, not otherwise apparent, was removed in 21; in 14 other patients unnecessary dissection was avoided when the new image confirmed that the surgical goals had been achieved. Additional time incurred with use of the PoleStar diminished with increasing experience. Testing of the navigation tool confirmed its accuracy to be comparable to other "frameless stereotactic" devices, on the order of 2 mm; likewise accuracy was increased by using T1 weighted imaging with thin slices. The introduction of new, faster imaging sequences and targeting tools has helped to make the PoleStar N-10 a routine tool for intracranial surgical navigation, with the added benefit of updated imaging during the procedure.
机译:介绍了用于术中磁共振成像(iMRI)和集成手术导航的紧凑型设备的使用经验。该系统商业上称为PoleStar N-10(Odin Medical Technologies,Yokne'am,以色列),包括一个0.12特斯拉永磁体,该永磁体在常规OR表下对接。包括一个基于被动红外的光学导航系统,该设备可以由正在运行的神经外科医生和工作人员进行控制。我们已经使用PoleStar N-10对93名患者进行了手术。诊断包括神经胶质瘤27例,垂体腺瘤19例,脑膜瘤15例,其他40例。35例患者的颅底有病变,其中8例有后颅窝肿块。 51%的手术受成像影响。在21处切除了其他未见的其他病变;当新图像确认已达到手术目标时,在其他14位患者中避免了不必要的解剖。随着经验的增加,使用PoleStar所花费的额外时间减少了。导航工具的测试证实其准确性可与其他“无框立体定向”设备相媲美,约为2毫米。同样,通过使用带有薄片的T1加权成像可以提高准确性。新的,更快的成像序列和靶向工具的引入使PoleStar N-10成为颅内手术导航的常规工具,并且在手术过程中具有更新的成像优势。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号