Assessment of MRI issues at 7 T for 28 implants and other objects

机译:在7岁处评估MRI问题,以进行28个植入物和其他物体

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVE. Metallic implants are currently a contraindication for volunteer subjects and patients referred for 7-T examinations because of concerns related to magnetic field interactions and MRI-related heating. Artifacts may also be problematic. Therefore, the purpose of this investigation was to evaluate these MRI issues for 28 implants and other objects in association MATERIALS AND METHODS. Tests were performed at 7 T using standardized procedures to evaluate magnetic field interactions (translational attraction and torque) for all 28 items. MRI-related heating and artifacts were assessed using spin-echo and gradient-echo pulse sequences, respectively, for two aneurysm clips located within a transmit-receive head radiofrequency coil. RESULTS. Eight of the 28 items showed magnetic field interactions at levels that could pose risks to human subjects. The two aneurysm clips exhibited heating, but the temperature rise did not exceed 1癈. Artifacts were dependent on the material and dimensions of each aneurysm clip. CONCLUSION. These findings show that certain implants and objects may be acceptable for human subjects undergoing MRI examinations at 7 T, whereas others may involve possible risks. This information has important implications for individuals referred for MRI examinations at 7 T.
机译:客观的。由于与磁场相互作用和MRI相关的加热相关的疑虑,金属植入物目前是志愿者受试者和患者提到的患者的禁忌症。文物也可能是有问题的。因此,这项调查的目的是评估28种植入物和其他物体中的这些MRI问题,并在结合材料和方法中。测试在7张中使用标准化程序进行,以评估所有28项的磁场相互作用(平移吸引力和扭矩)。使用旋转回波和梯度回波脉冲序列评估MRI相关的加热和伪像,用于位于发射接收头射频线圈内的两个动脉瘤夹子。结果。 28项中的八种在可能对人类受试者构成风险的水平上显示磁场相互作用。两个动脉瘤夹表现出加热,但温度升高不超过1‰。伪影依赖于每个动脉瘤夹的材料和尺寸。结论。这些发现表明,某些植入物和物体可能是在7吨接受MRI检查的人受试者的情况下可接受的,而其他人可能涉及可能的风险。此信息对7吨提到MRI考试的个人具有重要意义。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号