Rapid battery depletion and loss of therapy due to a short circuit in bipolar DBS for essential tremor

机译:由于基本震颤的双极DBS的短路,快速的电池耗尽和治疗丧失

获取原文
获取原文并翻译 | 示例
           

摘要

Technical dysfunctions have been reported reducing efficacy of deep brain stimulation (DBS). Here, we report on an essential-tremor patient in whom a short circuit in bipolar DBS resulted not only in unilateral loss of therapy but also in high current flow and thereby rapid decline of the impulse-generator battery voltage from 2.83 V a week before the event to 2.54 V, indicating the need for an impulse-generator replacement. Immediate re-programming restored therapeutic efficacy. Moreover, the reduction in current flow allowed the battery voltage to recover without immediate surgical intervention to 2.81 V a week later.
机译:据报道,技术功能障碍降低了深脑刺激(DBS)的疗效。 在这里,我们报告了双极DBS中短路的基本震颤患者,这不仅导致了单侧丧失疗法,而且导致高电流损失,从而在每周前2.83V的脉冲发电机电池电压迅速下降 事件到2.54 V,表示需要更换冲动发电机。 立即重新编程恢复治疗疗效。 此外,电流的减小允许电池电压在没有立即外科干预到每周2.81V的情况下恢复。