首页> 外文期刊>Acta Neurochirurgica >Unstable impedance of a single electrode contact resulting in loss of DBS therapya case report
【24h】

Unstable impedance of a single electrode contact resulting in loss of DBS therapya case report

机译:单个电极接触的不稳定阻抗导致DBS Therapya病例报告的损失

获取原文
获取原文并翻译 | 示例
           

摘要

yy Open and short circuits of electrode contacts are important technical dysfunctions of DBS. Here, we report on another type of dysfunction restricted to a single electrode contact: impedance instability within regular absolute values. After 9-year subthalamic DBS, a Parkinson patient developed unilateral motor symptoms and intermittent dysaesthesia due to impedance instability of the active contact. DBS efficacy could be restored without surgical revision by activation of the neighboring contact. During 3-year-follow-up, impedances of the dysfunctional contact varied between 1 and 3k whereas the other three contacts remained stable. Impedance documentation is crucial to identify such dysfunctions.
机译:电极接触的yy开放和短路是DBS的重要技术功能障碍。 在这里,我们报告限于单电极接触的另一种功能障碍:常规绝对值内的阻抗不稳定性。 在9年的次粒子DBS后,帕金森患者由于积极接触的阻抗不稳定而产生单侧运动症状和间歇性消化不良。 可以通过激活邻近的联系,在没有手术修订的情况下恢复DBS功效。 在3年随访期间,功能障碍接触的阻抗在1到3K之间变化,而另外三个触点保持稳定。 阻抗文件至关重要,以确定这种功能障碍。