Ocular microcirculation measurement with laser speckle flowgraphy and optical coherence tomography angiography in glaucoma

机译:具有激光散斑流程和光学相干断层造影血管造影的眼睛微循环测量

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Abstract Purpose To compare glaucoma severity with ocular microcirculation, measured with either laser speckle flowgraphy ( LSFG ) or optical coherence tomography angiography ( OCTA ). Methods We retrospectively studied 82 eyes of 82 open‐angle glaucoma ( OAG ) patients who underwent same‐day LSFG and OCTA examinations, with 20 eyes of 20 healthy subjects as controls. In OCTA images, vessel density ( VD ) (%) was calculated in concentric regions (regions 1, 2 and 3: R1, R2 and R3, respectively) defined by 1.6‐, 3.2‐ and 3.6‐mm‐diameter circles around the optic nerve head ( ONH ). In R3, the large vessels were automatically masked to calculate pure capillary density (auto R3 VD ). LSFG ‐measured mean blur rate ( MBR ) was examined in the overall ONH ( MA ), vessel‐area ONH ( MV ) and tissue‐area ONH ( MT ). Results Auto R3 VD had high reproducibility (coefficient of variation: 1.65–3.88%) and accurately reflected manual R3 VD [mean bias: ?0.0087% (auto R3 VD – manual R3 VD )]. MA and MT decreased significantly with OAG severity, especially in the early OAG stages (control vs. mild: p ? = ? 0.003 and p ? ? 0.001, respectively). R1 VD did not change with severity. R2 and auto R3 VD decreased significantly with OAG severity, especially in the severe stages (R2, mild vs. severe: p ? = ? 0.008; auto R3 VD , mild vs. severe: p ? ? 0.001, moderate vs. severe: p ? = ? 0.028). Conclusion Optical coherence tomography angiography‐derived auto R3 VD is novel, reproducible and accurately reflects manual measurements. It is useful for differentiating moderate and advanced glaucoma, while LSFG ‐derived MT is useful for identifying early glaucoma.
机译:摘要目的,用于比较眼睛微循环的青光眼严重程度,用激光散斑流程(LSFG)或光学相干断层造影血管造影(OctA)测量。方法我们回顾性地研究了82只眼睛的82个开放式青光眼(OAG)患者的82只眼睛,患有同日LSFG和Octa检查的患者,20只眼睛为20只健康受试者作为对照。在Octa图像中,在光学附近的1.6-,3.2-和3.6mm的圆圈中,在同心区域(区域1,2和3:R1,R2和R3)中计算血管密度(VD)(%)(分别为3.6mm直径圆圈神经头(ONH)。在R3中,大容器自动掩盖以计算纯毛细管密度(Auto R3 VD)。在整个ONH(MA),血管区ONH(MV)和组织区域ONH(MT)中检查LSFG -MEASUED平均模糊速率(MBR)。结果Auto R3 VD具有高再现性(变异系数:1.65-3.88%)和准确地反映手动R3 VD [均值:0.0087%(Auto R3 VD-Manual R3 VD)]。 MA和MT随着OAG严重程度而显着降低,特别是在早期的OAG阶段(对照与温和:P?= 0.003和P≤0.0.001)。 R1 VD没有严重变化。 R2和Auto R3 VD随着阳光严重程度显着降低,特别是在严重阶段(R2,轻度与严重:P?= 0.008; Auto R3 Vd,Mild Vs.严重:p≤x≤0.001,中等与严重:p?= 0.028)。结论光学相干断层造影血管造影术血管造影自动R3 VD是新颖的,可重复的,准确地反映了手动测量。它可用于区分中度和高级的青光眼是有用的,而LSFG -Derived MT可用于鉴定早期的青光眼。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号