首页> 外文期刊>Current Eye Research >Effect of Syringe Design on the Accuracy and Precision of Intravitreal Injections of Anti-VEGF Agents
【24h】

Effect of Syringe Design on the Accuracy and Precision of Intravitreal Injections of Anti-VEGF Agents

机译:注射器设计对抗VEGF试剂玻璃体内注射精度和精度的影响

获取原文
获取原文并翻译 | 示例
       

摘要

Purpose: To evaluate the accuracy and precision of different syringe designs for intravitreal injection of anti-VEGF agents.Methods: Volume output was measured from three syringe designs1) 1.0 mL tuberculin syringe, 2) 1.0 mL syringe with low dead space plunger, and 3) 0.5 mL low-volume syringeto deliver 50 mu L of bevacizumab, ranibizumab, or aflibercept, each repeated four times by three different physicians for 108 total simulated injections. Volume output was calculated from difference in syringe weight before and after expelling the drug. Accuracy was determined by mean absolute percentage error (MAPE), and precision was measured by coefficient of variation (CV).Results: Volume output from all three syringes was significantly different from 50 mu L, with mean volumes of 58.0 5.7 mu L for the tuberculin syringe, 58.0 +/- 4.0 mu L for the low dead space syringe, and 55.5 +/- 5.1 mu L for the low-volume syringe (p < 0.00001 for all). The low-volume syringe was the most accurate (MAPE = 12.8 +/- 7.8% vs. 17.3 +/- 9.3% or 15.9 +/- 8.1%), and the low dead space syringe was the most reproducible (CV = 0.068 vs. 0.091 or 0.097). There was no significant difference in volume output between different anti-VEGF agents.Conclusions: Intravitreal injections of anti-VEGF agents using a 1.0 mL tuberculin syringe demonstrate poor accuracy and precision. A lower capacity syringe may improve accuracy, while a low dead space plunger may improve precision.
机译:目的:评价不同注射器设计的抗VEGF剂的精度和精度。方法:从三个注射器设计中测量体积输出1)1.0mL结核蛋白注射器,2)1.0ml注射器,具有低死空间柱塞,3 )0.5ml低卷syringeo递送50 mu l的贝伐单抗,ranibizumab或aflibercept,每次重复三次不同的医生,持续108个总模拟注射。在排出药物之前和之后的注射器重量差异计算体积输出。通过平均绝对百分比误差(mape)确定精度,并且通过变异系数(cv)测量精度。结果:来自所有三个注射器的体积输出与50 mu l显着不同,平均5.0 5.7 mu l用于低死空间注射器的肺结核素注射器,58.0 +/- 4.0 mu l为低容量注射器55.5 +/- 5.1 mu l(全部为p <0.00001)。低批量注射器是最准确的(Mape = 12.8 +/- 7.8%与17.3 +/- 9.3%或15.9 +/- 8.1%),低死空间注射器是最重复的(CV = 0.068 Vs 。0.091或0.097)。不同抗VEGF试剂之间的体积输出差异无显着差异较低的容量注射器可以提高精度,而低死空间柱塞可以提高精度。

著录项

  • 来源
    《Current Eye Research》 |2017年第7期|共5页
  • 作者单位

    Univ Calif Davis Ctr Eye Dept Ophthalmol &

    Vis Sci 4860 Y St Suite 2400 Sacramento CA 95817 USA;

    Univ Calif Davis Ctr Eye Dept Ophthalmol &

    Vis Sci 4860 Y St Suite 2400 Sacramento CA 95817 USA;

    Univ Calif Davis Ctr Eye Dept Ophthalmol &

    Vis Sci 4860 Y St Suite 2400 Sacramento CA 95817 USA;

    Univ Calif Davis Ctr Eye Dept Ophthalmol &

    Vis Sci 4860 Y St Suite 2400 Sacramento CA 95817 USA;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类 眼科学;
  • 关键词

    Accuracy; anti-VEGF; intravitreal injection; precision; syringe;

    机译:准确性;抗VEGF;玻璃体内注射;精确;注射器;