首页> 外文期刊>Acta orthopaedica. >Exposure to chromium, cobalt and molybdenum from metal-on-metal total hip replacement and hip resurfacing arthroplasty.
【24h】

Exposure to chromium, cobalt and molybdenum from metal-on-metal total hip replacement and hip resurfacing arthroplasty.

机译:金属对金属的全髋关节置换和髋关节表面置换术暴露于铬,钴和钼。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: All metal implants--and metal-on-metal bearings in particular--corrode and cause a release of metal ions. Because cobalt and chromium have been shown to be carcinogenic and mutagenic in human and animal models, systemic toxicity and cancer risk are considered to be possible disadvantages of the metal-on-metal articulation. This study was designed to investigate the serum concentration profiles of chromium, cobalt and molybdenum after implantation of a Birmingham hip resurfacing arthroplasty (BHR) and a cementless total hip replacement with a 28-mm Metasul articulation (MTHR), over the first 2 years after implantation. METHODS: We analyzed profiles of metal ion serum levels in 111 patients implanted with a BHR, in 74 patients implanted with an MTHR, and in 130 implant-free probands control subjects using atomic absorption spectrophotometry. RESULTS: Chromium and cobalt concentrations (in microg/L) of all BHR and MTHR patients differed significantly from those of control subjects (chromium: <0.25; cobalt: 0.25). The median chromium and cobalt concentrations in BHR patients had increased to 5.1 and 4.3 microg/L 2 years after surgery. Concentrations in BHR patient exceeded those in the unilateral MTHR patients. Molybdenum serum concentrations hardly changed over time in either group and were not significantly different from the concentrations seen in the control subjects. INTERPRETATION: During the first 2 years after surgery, the Birmingham hip resurfacing arthroplasty leads to a significantly greater increase in serum chromium and cobalt levels than the 28-mm metal-on-metal MTHR. Observation of patients over a longer period will be necessary in order to evaluate any chronic adverse effects to the system due to elevated chromium and cobalt serum concentrations.
机译:背景:所有的金属植入物,特别是金属对金属的轴承,都会腐蚀并导致金属离子的释放。由于在人类和动物模型中钴和铬已被证明具有致癌性和致突变性,因此系统毒性和癌症风险被认为是金属对金属结合的可能缺点。这项研究旨在调查伯明翰髋关节置换术(BHR)植入和28 mm Metasul关节置换术(MTHR)的非骨水泥性全髋置换术后,头2年的铬,钴和钼的血清浓度曲线。植入。方法:我们使用原子吸收分光光度法分析了111例BHR植入患者,74例MTHR植入患者和130例无植入先证者对照受试者的金属离子血清水平。结果:所有BHR和MTHR患者的铬和钴浓度(以微克/升计)与对照组相比有显着差异(铬:<0.25;钴:0.25)。术后2年,BHR患者的铬和钴的中位数浓度已增加到5.1和4.3 microg / L。 BHR患者的浓度超过了单侧MTHR患者的浓度。两组中的钼血清浓度几乎不会随时间变化,并且与对照组的浓度没有显着差异。解释:在手术后的前2年中,伯明翰髋关节表面置换术比28毫米金属对金属MTHR导致血清铬和钴水平的明显增加。为了评估由于铬和钴血清浓度升高而对系统造成的任何慢性不利影响,有必要对患者进行更长时间的观察。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号