首页> 外文期刊>Acta orthopaedica. >No adverse effects of submelt-annealed highly crosslinked polyethylene in cemented cups.
【24h】

No adverse effects of submelt-annealed highly crosslinked polyethylene in cemented cups.

机译:水泥杯中的亚熔融退火高交联聚乙烯无不利影响。

获取原文
获取原文并翻译 | 示例
           

摘要

Highly crosslinked polyethylene (PE) is in standard use worldwide. Differences in the crosslinking procedure may affect the clinical performance. Experimenatal data from retrieved cups have shown free radicals and excessive wear of annealed highly crosslinked PE. We have previously reported low wear and good clinical performance after 6 years with this implant, and now report on the 10-year results.In 8 patients, we measured wear of annealed highly crosslinked PE prospectively with radiostereometry after 10 years. Activity was assessed by UCLA activity score and a specifically designed activity score. Conventional radiographs were evaluated for osteolysis and clinical outcome by the Harris hip score (HHS).The mean (95% CI) proximal head penetration for highly crosslinked PE after 10 years was 0.07 (-0.015 to 0.153) mm, and the 3D wear was 0.2 (0.026 to 0.36) mm. Without creep, proximal head penetration was 0.02 (-0.026 to 0.066) mm and for 3D penetration was 0.016 (-0.47 to 0.08) mm. This represents an annual proximal wear of less than 2 μm. All cups were clinically and radiographically stable but showed a tendency of increased rotation after 5 years.Wear for annealed highly crosslinked PE is extremely low up to 10 years. Free radicals do not affect mechanical performance or lead to clinically adverse effects. Creep stops after the first 6 months after implantation. Highly crosslinked PE is a true competitor of hard-on-hard bearings.
机译:高度交联的聚乙烯(PE)在世界范围内是标准用途。交联程序的差异可能会影响临床表现。取回的杯子的实验数据表明,自由基和过度磨损的退火高度交联的PE。我们以前曾报道该植入物在6年后具有较低的磨损和良好的临床表现,现在报道了10年的结果。在8例患者中,我们在10年后通过放射立体测量法对退火的高度交联的PE的磨损进行了前瞻性测量。通过UCLA活动评分和专门设计的活动评分来评估活动。传统的X线照片通过Harris髋关节评分(HHS)评估了骨溶解和临床结果。高度交联的PE在10年后的平均(95%CI)近端头穿透为0.07(-0.015至0.153)mm,3D磨损为0.2(0.026至0.36)毫米。如果没有蠕变,则近端头穿透力为0.02(-0.026至0.066)mm,而3D穿透力为0.016(-0.47至0.08)mm。这表示年度近端磨损小于2μm。所有杯在临床和放射学上均稳定,但在5年后显示出旋转增加的趋势。退火的高度交联PE的磨损在10年内极低。自由基不会影响机械性能或导致临床不良反应。植入后的前6个月,蠕变停止。高度交联的PE是硬对硬轴承的真正竞争对手。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号