Reduced radiation dose and improved image quality using a mini mobile digital imaging system in a neonatal intensive care unit

机译:使用迷你移动数字成像系统在新生儿重症监护室中减少辐射剂量和改进的图像质量

获取原文
获取原文并翻译 | 示例
       

摘要

This study was aimed to assess the radiation dose and image quality of a mini-mobile digital imaging (mini-DI) system for neonatal chest radiography and compared to conventional digital radiography (DR). A total of 64 neonates were examined and anatomical landmarks were assessed. The entrance surface dose of mini DI and conventional DR was 26.64 +/- 0.15 mu Gy and 49.11 +/- 1.46 mu Gy, respectively (p <0.001). The mean SNR values for mini DI and DR were 233.2 +/- 5.1 and 31.6 +/- 1.2, and 10% MTF values were 131 and 161 pm. A newly developed mini DI is capable of preserving the diagnostic information with dose reduction in neonates under intensive care. (C) 2016 Elsevier Inc. All rights reserved.