Multicenter, multivendor phantom study to validate proton density fat fraction and T2* values calculated using vendor-provided 6-point DIXON methods

机译:多中心,多选型幻影研究以验证使用供应商提供的6点达克松方法计算质子密度脂肪分数和T2 *值

获取原文
获取原文并翻译 | 示例

摘要

PurposeTo evaluate the reproducibility of proton density fat fraction (PDFF) and T2* in a fat-water phantom on three different 3?T MRI systems using 6-point DIXON methods. MethodsA phantom which included varying fat volume percentages (true fat fraction [FF]) was scanned by three 3?T MR machines, and PDFF and T2* were measured. ResultsThe mean difference between true FF and PDFF was small in all vendors (?2.11% to 0.41%). However, the difference ratio for T2* values was large among vendors (1.79 to 3.36). ConclusionsTrue FF and PDFF were consistent across vendors; however, T2* varied greatly.
机译:Purposeto使用6点Dixon方法评估三种不同3·T MRI系统的质子密度脂肪分数(PDFF)和T2 *的再现性。 方法包括不同脂肪体积百分比(真正的脂肪级分)(真正的脂肪级分)扫描三个3?T mR机床,测量PDFF和T2 *。 结果FF和PDFF之间的平均差异在所有供应商中都很小(?2.11%至0.41%)。 然而,供应商之间的T2 *值的差值(1.79至3.36)。 结案,FF和PDFF在供应商之间一致; 然而,T2 *变化大。