The objective of this study was to comparatively evaluate three commercial whole-blood platelet function analyzer systems: Platelet Function Analyzer-200 (PFA; Siemens Canada, Mississauga, Ontario, Canada), Multiplate analyzer (MP; Roche Diagnostics International Ltd., Rotkreuz, Switzerland), and Plateletworks Combo-25 kit (PLW; Helena Laboratories, Beaumont, TX, USA). Venipuncture was performed on 160 patients who visited a department of cardiology. Pairwise agreement among the three platelet function assays was assessed using Cohen's kappa coefficient and percent agreement within the reference limit. Kappa values with the same agonists were poor between PFA-collagen (COL; agonist)/adenosine diphosphate (ADP) and MP-ADP (-0.147), PFA-COL/ADP and PLW-ADP (0.089), MP-ADP and PLW-ADP (0.039), PFA-COL/ADP and MP-COL (-0.039), and between PFA-COL/ADP and PLW-COL (-0.067). Nonetheless, kappa values for the same assay principle with a different agonist were slightly higher between PFA-COL/ADP and PFA-COL/EPI (0.352), MP-ADP and MP-COL (0.235), and between PLW-ADP and PLW-COL (0.247). The range of percent agreement values was 38.7% to 73.8%. Therefore, various measurements of platelet function by more than one method were needed to obtain a reliable interpretation of platelet function considering low kappa coefficient and modest percent agreement rates among 3 different platelet function tests.
机译:本研究的目的是相对评价三种商业全血小板函数分析仪系统:血小板功能分析仪-200(PFA;西门子加拿大,Mississauga,安大略省,加拿大),多板分析仪(MP; Roche Diagnostics International Ltd.,Rotkreuz,Rotkreuz,瑞士)和血小板组合-25套件(PLW; Helena Laboratories,Beaumont,TX,USA)。静脉穿刺对160名患者进行了访问过的心脏病学。使用Cohen的Kappa系数和参考限制内的百分比评估三个血小板函数测定中的成对协议。具有相同激动剂的Kappa值在PFA-Collagen(COL;激动剂)/腺苷二磷酸(ADP)和MP-ADP(-0.147),PFA-COL / ADP和PLW-ADP(0.089),MP-ADP和PLW之间差-ADP(0.039),PFA-COL / ADP和MP-COL(-0.039)以及PFA-COL / ADP和PLW-COL(-0.067)。尽管如此,在PFA-COL / ADP和PFA-COL / EPI(0.352),MP-ADP和MP-COL(0.235)之间,以及PLW-ADP和PLW之间的相同测定原理的Kappa值略高于略高-col(0.247)。协议值百分比为38.7%至73.8%。因此,需要多种方法的各种测量血小板函数,以获得考虑低κ系数和3种不同血小板函数测试之间的血小板函数的可靠解释。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号