首页> 外文期刊>Clinical chemistry and laboratory medicine: CCLM >The effect of hematocrit on the results of measurements using glycose meters based on different techniques
【24h】

The effect of hematocrit on the results of measurements using glycose meters based on different techniques

机译:血细胞比容对使用不同技术的糖仪测量结果的影响

获取原文
获取原文并翻译 | 示例

摘要

Background: The aim of the study was to evaluate the effect of hematocrit (HCT) on glucose meter assays based on different measurement techniques. Methods: This paper studied glucose meters utilizing the glucose dehydrogenase reaction and four measurement techniques: colorimetry (HemoCue), reflectometry (Accu-Chek Active), amperometry (Optium Xido) and coulometry (Optium Omega). The EDTA venous blood samples HCT were modified by adding or removing defined aliquots of plasma. Glucose concentration was measured using each meter in 27 batches of blood samples, with HCT ranging from 20% to 60% in 10% increments. The data were analyzed using repeated measures models and a linear random effects model. Results: A significant relationship between HCT and glucose reading in all meters was found and, for all meters except Optium Xido, there was a significant modification of this relationship by glucose level. The relative decrease in glucose concentration per 1% increase of the HCT value varied from 0.30% for Optium Omega in samples with glucose concentrations <5.55 mmol/L to 1.37% for Optium Xido in the same stratum (p<0.0001). The 5% glucose meter error (the ADA recommendation) was reached in the <5.55 mmol/L stratum after HCT change by 3.9%-16.7%. Conclusions: There is a significant continuous effect of HCT on measurement accuracy of glucose meters across its wide range of values and glucose concentrations. The most sensitive to the HCT interference was the system utilizing amperometric technique (Optium Xido) followed by the one with reflectometric technique (Accu-Chek Active), while the systems with the coulometric technique (Optium Omega) or colorimetric measurements in whole blood haemolysate (HemoCue) were less sensitive.
机译:背景:该研究的目的是评估血细胞比容(HCT)对基于不同测量技术的葡萄糖表测定的影响。方法:本文研究了利用葡萄糖脱氢酶反应和四种测量技术的葡萄糖计:比色法(血管),反射测量(Accu-Chek Active),安培(XIDO)和库仑(Optium Omega)。通过加入或除去定义的等离子体来改变EDTA静脉血液样品HCT。使用27批血液样品中的每个仪表测量葡萄糖浓度,随着10%的增量,HCT为20%至60%。使用重复测量模型和线性随机效果模型进行分析数据。结果:在所有仪表之间找到了HCT与葡萄糖阅读的重要关系,并且对于XIDO除外的所有米,通过葡萄糖水平对这种关系进行了重大修改。 HCT值增加的葡萄糖浓度的相对降低从葡萄糖浓度的样品中的戊uMega的0.30%变化,在同一地层​​中的XIDO的葡萄糖浓度<5.55mmol / L至1.37%(P <0.0001)。在HCT变化后,在<5.55mmol / L层中达到5%葡萄糖仪误差(ADA推荐)达到3.9%-16.7%。结论:HCT对其广泛的价值和葡萄糖浓度的葡萄糖计的测量精度显着持续效果。对HGT干扰最敏感的是利用安培测量技术(Optium Xido),然后是具有反射测量技术(Accu-Chek Active)的系统,而具有Coulometric技术(Optium Omega)的系统或全血氧态的比色测量( Hemocue)敏感。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号