A 62-year old male underwent treatment of a bifurcation lesion in the left anterior descending coronary artery using provisional stenting with a jailed wire technique. Severe longitudinal stent deformation and unravelment of the stent part proximal of the bifurcation occurred when we tried to pull the jailed wire from the side branch. The described case poses a caution on the use of this specific stent(-design) in bifurcation lesions.
机译:一种62岁的男性在左前期下降冠状动脉中进行分叉病变的疗法,使用临时支架与入狱丝技术进行临时支架。 当我们试图从侧面分支中拉动导丝时,在分叉的近端的严重纵向支架变形和解析的支架部分发生。 所描述的情况在分叉病变中使用这种特定支架(-design)致致谨慎。