A comparison of patient position displacements from body surface laser scanning and cone beam CT bone registrations for radiotherapy of pelvic targets

机译:体表激光扫描和锥束CT骨定位对骨盆靶标放射治疗的患者位置位移的比较

获取原文
获取原文并翻译 | 示例

摘要

Background and purpose. Optical surface detection has attractive features as a mean in radiotherapy for patient positioning tasks such as set-up, monitoring and gating. To aid in hitting radiotherapy targets the correlation between detected surface displacements and internal structure displacements is crucial. In this study, we compare set-up displacements derived from a body surface laser scanning (BSLS) system to displacements derived from bone registrations with a cone beam computed tomography (CBCT) system in order to quantify the accuracy and applicability of BSLS for fractionated treatments in the pelvic region. Material and methods. Displacements from concurrent BSLS and CBCT registrations were compared for 40 patients treated in the pelvic region for a total of 170 set-ups. Surface data captured by BSLS at the first treatment fraction (BSLSref) was used as main reference for the BSLS system, while bony structures from the planning CT were used as a reference for the CBCT method. As comparison, the patient outline extracted from the planning CT was used as BSLS reference (CTref). The displacements detected by the CBCT system (skin-marks-only) was also used for comparison. Results. The mean differences (+/- 1 SD) between the BSLS and CBCT displacements were -0.01 (+0.17) cm, 0.00 (+0.21) cm and 0.01 (+0.17) cm in the lateral, longitudinal and vertical directions, respectively. The median length of the difference was 0.26 cm (0.24-0.29 cm, 95% CI). The median of the difference between CBCT and BSLS displacements based on CTref was 0.37 cm (0.30-0.39 cm) and the median for skin-marks-only was 0.38 cm (0.34-0.42 cm). Conclusions. The BSLS system is a good supplement to the CBCT system for accurate set-up for fractions when no CBCT is deemed necessary for pelvic targets. Inter-fractional skin movement in relation to bone was estimated to be 0.2 cm in the lateral (X), longitudinal (Y) and vertical direction (Z), respectively.
机译:背景和目的。光学表面检测具有很强的吸引力,作为放射治疗中诸如定位,监视和门控等患者定位任务的一种手段。为了帮助达到放射治疗目标,检测到的表面位移与内部结构位移之间的相关性至关重要。在这项研究中,我们将体表激光扫描(BSLS)系统产生的位移与通过锥束计算机断层扫描(CBCT)系统从骨配准产生的位移进行比较,以量化BSLS在分级治疗中的准确性和适用性在骨盆区域材料与方法。比较了同时进行的BSLS和CBCT注册的位移,对在骨盆区域接受治疗的40位患者进行了总共170项设置。 BSLS在第一个处理分数(BSLSref)处捕获的表面数据被用作BSLS系统的主要参考,而计划CT的骨结构被用作CBCT方法的参考。作为比较,将从计划CT中提取的患者轮廓用作BSLS参考(CTref)。由CBCT系统检测到的位移(仅皮肤标记)也用于比较。结果。 BSLS和CBCT位移之间的平均差异(+/- 1 SD)在横向,纵向和垂直方向分别为-0.01(+0.17)cm,0.00(+0.21)cm和0.01(+0.17)cm。差异的中值长度为0.26厘米(0.24-0.29厘米,95%CI)。基于CTref的CBCT和BSLS位移之间的差异的中位数为0.37厘米(0.30-0.39厘米),仅有皮肤痕迹的中位数为0.38厘米(0.34-0.42厘米)。结论当认为骨盆目标不需要CBCT时,BSLS系统是CBCT系统的良好补充,可精确设置馏分。相对于骨骼的小部分皮肤间运动在横向(X),纵向(Y)和垂直方向(Z)分别估计为0.2 cm。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号