首页> 外文期刊>Cardiovascular and Interventional Radiology: A Journal of Imaging in Diagnosis and Treatment >Broken Wings: First Reported Cases of Fractured Angel (R) Catheters, a Temporary Combined Femoral Venous Catheter and Retrievable IVC Filter Device
【24h】

Broken Wings: First Reported Cases of Fractured Angel (R) Catheters, a Temporary Combined Femoral Venous Catheter and Retrievable IVC Filter Device

机译:破碎的翅膀:首次报道的裂缝天使(R)导管,临时组合股骨静脉导管和可检索的IVC过滤装置

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

There has been increasing use of a novel combined femoral venous sheath, catheter and retrievable self-expanding and collapsible diamond-shaped IVC filter (Angel(A (R)) Catheter, BiO2 Medical), in severely injured patients who cannot receive anticoagulation. As the filter is not detached from the catheter/sheath, it should be easily retrieved. Outcomes included in large registries demonstrate a high safety profile and a 100% retrieval rate. However, at our institution-a Level 1 major UK trauma centre with 4 years of substantial experience in using this device-we've encountered three cases of device fracture and subsequent complicated retrieval dating from Dec 2016 to March 2017. To the best of the authors' knowledge, we describe the first documented case series of fractured Angel(A (R)) Catheters and their retrieval.
机译:在不可接受抗凝的严重受伤的患者中,一直越来越多地利用了一种新颖的股骨静脉护套,导管和可检索的自膨胀和可折叠的钻石形IVC过滤器(天使(A(R))导管,Bio2医疗)。 由于过滤器未与导管/护套拆卸,因此应该容易地检索。 大型注册表中包含的结果表明了高安全性和100%的检索率。 但是,在我们的机构 - 一级级别的英国Trauma中心,使用该设备4年的大量经验 - 我们已经遇到了三种设备骨折和随后复杂的复杂检索从2016年12月至2017年3月。 作者的知识,我们描述了第一系列裂缝天使(A(R))导管及其检索。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号