Low radiation dose in computed tomography: the role of iodine

机译:计算断层扫描中的低辐射剂量:碘的作用

获取原文
获取原文并翻译 | 示例
           

摘要

Recent approaches to reducing radiation exposure during CT examinations typically utilize automated dose modulation strategies on the basis of lower tube voltage combined with iterative reconstruction and other dose-saving techniques. Less clearly appreciated is the potentially substantial role that iodinated contrast media (CM) can play in low-radiation-dose CT examinations. Herein we discuss the role of iodinated CM in low-radiation-dose examinations and describe approaches for the optimization of CM administration protocols to further reduce radiation dose and/or CM dose while maintaining image quality for accurate diagnosis. Similar to the higher iodine attenuation obtained at low-tube-voltage settings, high-iodine-signal protocols may permit radiation dose reduction by permitting a lowering of mAs while maintaining the signal-to-noise ratio. This is particularly feasible in first pass examinations where high iodine signal can be achieved by injecting iodine more rapidly. The combination of low kV and IR can also be used to reduce the iodine dose. Here, in optimum contrast injection protocols, the volume of CM administered rather than the iodine concentration should be reduced, since with high-iodine-concentration CM further reductions of iodine dose are achievable for modern first pass examinations. Moreover, higher concentrations of CM more readily allow reductions of both flow rate and volume, thereby improving the tolerability of contrast administration.
机译:最近在CT检查期间减少辐射暴露的方法通常根据较低管电压与迭代重建和其他剂量节省技术相结合自动化剂量调制策略。较少明确地理解是碘化造影剂(CM)可以在低辐射剂量CT检查中发挥的潜在实质性作用。在此我们讨论碘化CM在低辐射剂量检查中的作用,并描述了用于优化CM给药方案的方法,以进一步减少辐射剂量和/或CM剂量,同时保持图像质量以进行准确诊断。类似于在低管电压设置下获得的较高的碘衰减,高碘信号协议可以通过在保持信噪比的同时降低MAS来允许辐射剂量降低。这在第一通过检查中特别可行,其中通过更快地注入碘可以实现高碘信号。低kV和IR的组合也可用于减少碘剂量。这里,在最佳对比注射方案中,施用的CM体积而不是碘浓度应减少,因为对于高碘浓度的CM,可进一步减少碘剂量,可用于现代首次通过检查。此外,更高浓度的Cm更容易允许减少流速和体积,从而提高对比度给药的可耐受性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号