Applying the Allan Variance to the Analysis of Errors of the Expiration Volume Measurement Channel of Mechanical Lung Ventilators

机译:将Allan差异应用于机械肺呼吸机的呼气量测量通道误差分析

获取原文
获取原文并翻译 | 示例
           

摘要

The expiration volume measurement channel of a mechanical lung ventilator operating at a low rate of gas mixture flow is considered. The channel components are described in detail. The importance of accurate measurement of gas flow at low rates of mechanical ventilation (especially in neonatal medicine) is substantiated. The Allan variance is suggested as a statistical tool for analyzing the measurement channel errors. The obtained Allan variances, their approximations as sums of exponential functions, and estimates of power spectrum density coefficients of zero-signal noise are presented. The obtained estimates of the main components of the channel error are compared with the parameters of the electronic components used in the measurements. Preliminary recommendations for reduction of errors of the volume measurement channel are given.
机译:考虑了以低气体混合物流动操作的机械肺气通道的到期体积测量通道。 详细描述信道组件。 确保了低速率的机械通风(特别是新生儿)低速率测量气流的重要性。 建议Allan方差作为分析测量信道错误的统计工具。 所获得的Allan差异,它们的近似作为指数函数的和的总和,以及零信号噪声的功率谱密度系数的估计。 将所得信道误差的主要组件的估计与测量中使用的电子元件的参数进行比较。 给出了减少体积测量通道错误的初步建议。