【24h】

Disposable tyrosinase-peroxidase bi-enzyme sensor for amperometric detection of phenols

机译:一次性酪氨酸酶-过氧化物酶双酶传感器用于苯酚的安培检测

获取原文
获取原文并翻译 | 示例

摘要

A new disposable amperometric bi-enzyme sensor system for detecting phenols has been developed. The phenol sensor developed uses horseradish peroxidase modified screen-printed carbon electrodes (HRP-SPCEs) coupled with immobilized tyrosinase prepared using poly(carbamoylsulfonate) (PCs) hydrogels or a poly(vinyl alcohol) bearing styrylpyridinium groups (PVA-SbQ) matrix. Optimization of the experimental parameters has been performed with regard to buffer composition, pH, operating potential and storage stability. A co-operative reaction involving tyrosinase and HRP occurs at a potential of -50 mV versus Ag/AgCl without the requirement for addition of extraneous H2O2, thus, resulting in a very simple and efficient system. Comparison of the electrode responses with the 4-aminoantipyrine standard method for phenol sample analysis indicated the feasibility of the disposable sensor system for sensitive 'in-field' determination of phenols. The most sensitive system was the tyrosinase immobilized HRP-SPCE using PCs, which displayed detection limits for phenolic compounds in the lower nanomolar range e.g. 2.5 nM phenol, 10 nM catechol and 5 nM p-cresol. (C) 2002 Elsevier Science B.V. All rights reserved. [References: 34]
机译:已经开发出一种用于检测苯酚的新型一次性安培双酶传感器系统。开发的酚传感器使用辣根过氧化物酶修饰的丝网印刷碳电极(HRP-SPCE)结合固定化酪氨酸酶,该酪氨酸酶是使用聚(氨甲酰基磺酸盐)(PC)水凝胶或带有苯乙烯吡啶鎓基团(PVA-SbQ)的聚乙烯醇制备的。关于缓冲液组成,pH,操作电位和储存稳定性,已经对实验参数进行了优化。相对于Ag / AgCl,涉及酪氨酸酶和HRP的协同反应发生在-50 mV的电位,而不需要添加多余的H2O2,因此,系统非常简单有效。电极响应与4-氨基安替比林标准方法进行酚样品分析的比较表明,一次性传感器系统可用于灵敏的“现场”测定酚。最灵敏的系统是使用PC固定化酪氨酸酶的HRP-SPCE,它在较低的纳摩尔范围内显示了对酚类化合物的检测限,例如2.5 nM苯酚,10 nM邻苯二酚和5 nM对甲酚。 (C)2002 Elsevier Science B.V.保留所有权利。 [参考:34]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号