摘要

Two new amperometric carbon paste enzyme electrodes were developed for xanthine determination. 1,4-benzoquinone and poly(vinylferrocene) (PVF) were investigated as mediators. The parameters affecting the analytical performance of the enzyme electrode have been investigated in detail and optimized for modified enzyme electrodes. 1,4-benzoquinone modified enzyme electrode (BQ-CPEE) exhibited linear response from 1.9 x 10~(-7)M to 5.5 x10~(-6) M and from 5.2 x 10~(-5) M to 8.2 x 10~(-4) M with a good detection limit of 1.0 x 10~(-7)M. The linear working range of the PVF modified enzyme electrode was between 1.9 x 10~(-7)-2.1 x 10~(-6) M, 1.9 x 10~(-6)-1.0 x 10~(-5) M and 1.1 x 10~(-4)-8.8 x 10~(-4) M with a detection limit of 1.0 x 10~(-7)M. Hypoxanthine response of the electrodes was also determined. Modified enzyme electrodes were used for xanthine determination in real samples and good recoveries were obtained.
机译:开发了两个用于安定黄嘌呤的新的安培型碳糊酶电极。研究了1,4-苯醌和聚乙烯基二茂铁(PVF)作为介体。已经详细研究了影响酶电极分析性能的参数,并针对修饰的酶电极进行了优化。 1,4-苯醌修饰的酶电极(BQ-CPEE)表现出从1.9 x 10〜(-7)M到5.5 x10〜(-6)M和从5.2 x 10〜(-5)M到8.2 x 10的线性响应〜(-4)M,检测限为1.0 x 10〜(-7)M。 PVF修饰酶电极的线性工作范围为1.9 x 10〜(-7)-2.1 x 10〜(-6)M,1.9 x 10〜(-6)-1.0 x 10〜(-5)M和1.1 x 10〜(-4)-8.8 x 10〜(-4)M,检测极限为1.0 x 10〜(-7)M。还确定了电极的次黄嘌呤响应。修饰的酶电极用于实际样品中的黄嘌呤测定,并获得了良好的回收率。