Effect of automatic tube current modulation on radiation dose and image quality for low tube voltage multidetector row CT angiography: phantom study.

机译:自动管电流调制对低管电压多探测器行CT血管造影的辐射剂量和图像质量的影响:幻像研究。

获取原文
获取原文并翻译 | 示例
           

摘要

RATIONALE AND OBJECTIVES: To evaluate the effect of automatic tube current modulation on radiation dose and image quality for low tube voltage computed tomography (CT) angiography. MATERIALS AND METHODS: An anthropomorphic phantom was scanned with a 64-section CT scanner using following tube voltages: 140 kVp (Protocol A), 120 kVp (Protocol B), 100 kVp (Protocol C), and 80 kVp (Protocol D). To achieve similar noise, combined z-axis and xy-axes automatic tube current modulation was applied. Effective dose (ED) for the four tube voltages was assessed. Three plastic vials filled with different concentrations of iodinated solution were placed on the phantom's abdomen to obtain attenuation measurements. The signal-to-noise ratio (SNR) was calculated and a figure of merit (FOM) for each iodinated solution was computed as SNR(2)/ED. RESULTS: The ED was kept similar for the four different tube voltages: (A) 5.4 mSv +/- 0.3, (B) 4.1 mSv +/- 0.6, (C) 3.9 mSv +/- 0.5, and (D) 4.2 mSv +/- 0.3 (P > .05). As the tube voltage decreased from 140 to 80 kVp, image noise was maintained (range, 13.8-14.9 HU) (P > .05). SNR increased as the tube voltage decreased, with an overall gain of 119% for the 80-kVp compared to the 140-kVp protocol (P < .05). The FOM results indicated that with a reduction of the tube voltage from 140 to 120, 100, and 80 kVp, at constant SNR, ED was reduced by a factor of 2.1, 3.3, and 5.1, respectively, (P < .001). CONCLUSIONS: As tube voltage decreases, automatic tube current modulation for CT angiography yields either a significant increase in image quality at constant radiation dose or a significant decrease in radiation dose at a constant image quality.
机译:理由和目的:评价低管电压计算机断层扫描(CT)血管造影的自动管电流调制对辐射剂量和图像质量的影响。材料与方法:使用64段CT扫描仪使用以下电子管电压扫描拟人模型:140 kVp(协议A),120 kVp(协议B),100 kVp(协议C)和80 kVp(协议D)。为了获得相似的噪声,应用了组合的z轴和xy轴自动管电流调制。评估了四个管电压的有效剂量(ED)。将三个装有不同浓度碘溶液的塑料小瓶放在体模的腹部,以进行衰减测量。计算信噪比(SNR),并将每种含碘溶液的品质因数(FOM)计算为SNR(2)/ ED。结果:对于四个不同的管电压,ED保持相似:(A)5.4 mSv +/- 0.3,(B)4.1 mSv +/- 0.6,(C)3.9 mSv +/- 0.5和(D)4.2 mSv +/- 0.3(P> .05)。当管电压从140 kVp降低到80 kVp时,图像噪声得以维持(范围13.8-14.9 HU)(P> 0.05)。信噪比随管电压的降低而增加,与140kVp协议相比,80kVp的总增益为119%(P <.05)。 FOM结果表明,当管电压从140降低到120、100和80 kVp时,在恒定SNR下,ED分别降低了2.1、3.3和5.1倍(P <.001)。结论:随着管电压的降低,用于CT血管造影的自动管电流调制可在恒定辐射剂量下显着提高图像质量,或在恒定图像质量下显着降低辐射剂量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号