There is currently a major issue with the calibration of conductivity meters used for high purity water: the lack of availability of a reference material or reference methods for low conductivity ranges (conductivity below 1 mu S cm(-1) at 25.0 degrees C, resistivity >1 M Omega cm at 25.0 degrees C). This paper describes the current status of conductivity measurements in high purity water. A new and improved approach, currently being investigated, should allow us to make the calibration of conductivity meters used for low conductivity ranges traceable to the SI.
机译:当前用于高纯水的电导率仪的校准存在一个主要问题:缺乏用于低电导率范围的参考材料或参考方法(在25.0摄氏度下电导率低于1μS cm(-1),电阻率在25.0摄氏度下> 1 MΩ厘米)。本文介绍了高纯水中电导率测量的当前状态。目前正在研究一种新的改进方法,这将使我们能够将用于低电导率范围的电导率仪的校准追溯到SI。