This paper describes a comparison calibration system applicable to meters for low electrolytic conductivity values. In this system, measurements are performed in a closed circuit with a flowing solution. This circuit contains a flow-through reference cell and a flow-through cell under calibration. Results of calibrations by substitution and by comparison at 3 mS/m (KCl aqueous solution) are compared. The result of a calibration by comparison at 0.3 mS/m is also reported.
机译:本文介绍了一种适用于电导率值较低的仪表的比较校准系统。在该系统中,使用流动的溶液在闭合回路中进行测量。该电路包含一个流通参考池和一个经过校准的流通池。比较了通过置换和以3 mS / m(KCl水溶液)进行比较的校准结果。还报告了通过比较以0.3 mS / m进行校准的结果。