Change in biomechanical parameters after flap creation using the Intralase femtosecond laser and subsequent excimer laser ablation.

机译:使用Intralase飞秒激光和随后的准分子激光消融术在皮瓣形成后生物力学参数的变化。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: This study aims to investigate the effect of both flap creation and laser ablation on corneal hysteresis (CH) and corneal resistance factor (CRF), as well as the inward applanation signal amplitude produced by the ocular response analyzer (ORA), immediately following each step of the LASIK procedure using the Intralase femtosecond laser for flap creation. METHODS: A total of 66 eyes of 35 subjects who underwent Intra-LASIK surgery were prospectively enrolled in this study. Changes in the CH, CRF, and amplitude of the first applanation peak (Peak 1) of the infrared signal were measured preoperatively, immediately after flap elevation and repositioning, as well as subsequent to laser ablation in all eyes. Repeated measures analysis of variance (ANOVA) was performed with a post-test of contrast variable profiles to investigate changes following each step of the procedure. RESULTS: The repeated measures ANOVA indicated a significant difference (p < 0.001) among the three time points for CH, CRF, and Peak 1. There was no significant change in CH (p = 0.20) or CRF (p = 0.57) after flap creation, but there was significant decrease in these parameters following laser ablation (p < 0.0001), when compared to values obtained pre-operatively. There was a significant decrease in Peak 1 both after flap creation (p < 0.0001) and after subsequent ablation (p < 0.0001), when compared to pre-operative ORA signal peak amplitudes. CONCLUSION: There is a well documented change in corneal biomechanical properties following a LASIK procedure. This study reveals that this change may be predominantly due to laser ablation. However, flap creation with the Intralase laser does produce a biomechanical consequence consistent with reduction of corneal stiffness, as measured by the reduced amplitude of Peak 1. CH and CRF do not fully characterize this change.
机译:目的:本研究旨在研究皮瓣产生和激光消融对眼角滞后(CH)和角膜抵抗因子(CRF)以及眼响应分析仪(ORA)产生的向内压平信号幅度的影响LASIK手术的每个步骤都使用Intralase飞秒激光进行皮瓣制作。方法:前瞻性研究了35例接受了LASIK手术的受试者的66只眼睛。术前,抬高皮瓣并重新定位后,以及在所有眼睛进行激光消融后,术前测量CH,CRF和红外信号第一个压平峰(峰1)的幅度变化。重复测量方差分析(ANOVA)进行对比变量配置文件的后测,以调查该过程每个步骤之后的变化。结果:重复测量方差分析表明CH,CRF和Peak 1在三个时间点之间存在显着差异(p <0.001)。皮瓣拍打后CH(p = 0.20)或CRF(p = 0.57)没有显着变化。与术前获得的值相比,激光烧蚀后这些参数显着降低(p <0.0001)。与术前ORA信号峰值幅度相比,皮瓣产生后(p <0.0001)和随后的消融(p <0.0001)之后,峰1均显着下降。结论:LASIK手术后,有充分记录的角膜生物力学特性变化。这项研究表明,这种变化可能主要是由于激光烧蚀引起的。但是,用Intralase激光制作皮瓣的确会产生与降低角膜僵硬度相一致的生物力学结果,这是由Peak 1幅度减小所测得的。CH和CRF不能完全表征这种变化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号