Performance evaluation of quantitative SPECT/CT using NEMA NU 2 PET methodology

机译:使用NEMA NU 2 PET方法的定量SPECT / CT的性能评估

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Although PET is routinely evaluated using NEMA NU2 as standard in the clinic, standard methodology for evaluating the performance of quantitative SPECT systems has not been established. In this study, the quantitative performance of the Symbia Intevo SPECT/CT was evaluated for two common isotopes (Tc-99m, Lu-177) and benchmarked against the performance of a PET/CT. A further aim was to demonstrate the utility of adapting NEMA NU2 PET measurements to SPECT. In addition, dead-time and resolution recovery were evaluated to provide more complete system evaluations. Spatial resolution of the SPECT system at 1 cm from the center in the transverse direction was 13.1 mm and 22.4 mm for Tc-99m and Lu-177 respectively, compared with 4.3 mm (F-18) and 5.8 mm (Ga-68) for PET. Sensitivity at the center of the FoV was 119 cps MBq(-1) and 48 cps MBq(-1) ((99)mTc,Lu-177) for SPECT and 9632 cps MBq(-1) and 8216 cps MBq(-1) (F-18,Ga-68) for PET. Scatter fraction was 0.25 and 0.36 ((99)mTc,Lu-77) for SPECT and 0.32 and 0.29 (F-18, Ga-68) for PET. Contrast recovery coefficient in the largest spheres was 0.79 and 0.65 (Tc-99m,Lu-177) for SPECT, 1.00 and 0.97 (F-18,Ga-68) for PET and the background variability was 2.7%, 6.5% (Tc-99m,Lu-177), 1.5% and 1.6% (F-18,Ga- 68), respectively. Partial volume effect was evaluated using the NEMA IQ phantom with six sphere inserts (diameter: 37 mm, 28 mm, 22 mm, 17 mm, 13 mm and 10 mm). Full contrast recovery was reached with the 17 mm for F-18, while SPECT did not reach full recovery for any sphere. Count rate losses were 2% for Tc-99m at 1 GBq and 11% for Lu-177 at 8.5 GBq which are well below the typical activities for clinical applications. We concluded NEMA NU2 methodology can be easily adapted to SPECT/CT as a routine quality assurance procedure in the clinic.
机译:虽然使用NEMA Nu2作为标准在诊所的标准中常规评估PET,但尚未建立用于评估定量SPECT系统性能的标准方法。在这项研究中,对Symbia Intevo Spect / CT的定量性能进行了评估了两个常见的同位素(TC-99M,LU-177)并与PET / CT的性能进行了基准测试。进一步的目的是证明将NEMA NU2 PET测量适用于SPECT的效用。此外,评估死区时间和分辨率恢复,以提供更完整的系统评估。与4.3mm(F-18)和5.8mm(GA-68)相比,TC-99M和LU-177分别在横向上的1厘米的SPECT系统的空间分辨率为13.1mm和22.4mm。宠物。 FOV中心的灵敏度为119 CPS MBQ(-1)和48个CPS MBQ(-1)((99)MTC,LU-177),SPECT和9632 CPS MBQ(-1)和8216 CPS MBQ(-1 PET的)(F-18,GA-68)。 SAPT分散馏分为0.25和0.36((99)MTC,LU-77),用于PET的0.32和0.29(F-18,GA-68)。对于SPECT,最大球体中的对比度回收系数为0.79和0.65(TC-99M,LU-177),宠物的1.00和0.97(F-18,GA-68),背景变异为2.7%,6.5%(TC- 99M,LU-177),1.5%和1.6%(F-18,GA-68)。使用带有六个球体插入件的NEMA IQ幻影评估部分体积效果(直径:37 mm,28 mm,22 mm,17 mm,13 mm和10mm)。对于F-18的17毫米,达到完全对比度恢复,而SPECT对任何球体没有达到全面恢复。计数率损失对于1 GBQ的TC-99M为2%,LU-177的11%在8.5 GBQ,远低于临床应用的典型活动。我们结束了Nema Nu2方法可以轻松适应Spect / CT作为诊所的常规质量保证程序。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号