首页> 外文期刊>Physics in medicine and biology. >Performance evaluation of the MOLECUBES beta-CUBE-a high spatial resolution and high sensitivity small animal PET scanner utilizing monolithic LYSO scintillation detectors
【24h】

Performance evaluation of the MOLECUBES beta-CUBE-a high spatial resolution and high sensitivity small animal PET scanner utilizing monolithic LYSO scintillation detectors

机译:分子β - 立方体的性能评估 - 一种利用单片Lyso闪烁探测器的高空间分辨率和高灵敏度小动物PET扫描仪

获取原文
获取原文并翻译 | 示例
       

摘要

The MOLECUBES beta-CUBE scanner is the newest amongst commercially available predinical PET scanners for dedicated small animal imaging. The scanner is compact, lightweight and utilizes a small footprint to facilitate bench-top imaging. It can be used individually, or in combination with the X-CUBE CT scanner, which provides the ability to perform all necessary PET data corrections and provide fully quantitative PET images. The PET detector comprises of an 8 mm thick monolithic LYSO scintillator read-out by an array of 3 mm x 3 mm Hamamatsu silicon photomultipliers. The monolithic scintillator provides the ability to measure depth-of-interaction which aids in the development of such a compact scanner. With a scanner diameter of 7.6 cm and axial length of 13 cm it is suitable for imaging both whole-body mice and rats. This paper presents the design and imaging performance of the beta-CUBE scanner. NEMA NU4-2008 characterization and a variety of phantom and animal imaging studies to demonstrate the quantitative imaging performance of the PET scanner are presented. Spatial resolution of 1 mm is measured with a filtered-back projection reconstruction algorithm at the center of the scanner and DOI measurement helps maintain the excellent spatial resolution over the entire imaging FOV. An absolute peak sensitivity of 12.4% is measured with a 255-765 keV energy window. The scanner demonstrates good count-rate performance, with a peak NEC of 300 kcps and 160 kcps measured with similar to 900 mu Ci in the NEMA mouse and rat phantoms, respectively. Imaging data with the NEMA image quality phantom and Micro Derenzo phantoms demonstrate the ability to achieve good image quality and accurate quantitative data. Image uniformity of 7.4% and spill-over ratio of 8% were measured. The superior spatial resolution, excellent energy resolution and sensitivity also provide superior contrast recovery, with similar to 70% recovery for the 2 mm rods. While current commercial predinical PET scanners have spatial resolution in the 1-2 mm range, the 1 mm 3 volumetric resolution presents significant improvement over current commercially available predinical PET scanners. In combination with the X-CUBE scanner it provides the ability to perform fully quantitative imaging with spatially co-registered high-resolution 3D PET-CT images.
机译:分子Beta-Cube扫描仪是商业上可用的宠物扫描仪中最新的专用小动物成像。扫描仪紧凑,轻巧,利用小占地面积来促进台式成像。它可以单独使用,或者与X-Cube CT扫描仪组合使用,其提供了执行所有必要的PET数据校正的能力并提供完全定量的PET图像。 PET检测器包括8毫米厚的整体式LYSO闪烁体,其阵列为3mm×3mm Hamamatsu硅光电倍增器。整体闪烁体提供测量相互作用的浓度的能力,这有助于这种紧凑型扫描仪的开发。扫描仪直径为7.6厘米,轴向长度为13厘米,适用于成像全体小鼠和大鼠。本文介绍了Beta-Cube扫描仪的设计和成像性能。提出了NEMA NU4-2008表征和各种幻像和动物成像研究,以证明PET扫描仪的定量成像性能。使用扫描仪中心的滤波后投影重建算法测量1mm的空间分辨率,并且DOI测量有助于在整个成像FOV上保持优异的空间分辨率。使用255-765 kev能量窗口测量12.4%的绝对峰值敏感性。扫描仪显示出良好的计数性能,峰值NEC为300kCPS,160kCP分别测量了NEMA小鼠和大鼠幻影的900 mu CI。与NEMA图像质量幻像和Micro Derenzo Phantoms的成像数据展示了实现良好的图像质量和准确的定量数据的能力。测量图像均匀性7.4%,溢出比率为8%。出色的空间分辨率,优异的能量分辨率和敏感性也提供了卓越的对比度恢复,而2 mm棒的回收率类似于70%。虽然目前的商业产品PET扫描仪在1-2毫米范围内具有空间分辨率,但是1mm 3容量分辨率在电流市售的产品PET扫描仪上具有显着的改进。结合X-Cube扫描仪,它提供了在空间共同注册的高分辨率3D PET-CT图像中执行完全定量成像的能力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号