首页> 外文期刊>Spine >The Crooked Rod Sign A New Radiological Sign to Detect Deformed Threads in the Distraction Mechanism of Magnetically Controlled Growing Rods and a Mode of Distraction Failure
【24h】

The Crooked Rod Sign A New Radiological Sign to Detect Deformed Threads in the Distraction Mechanism of Magnetically Controlled Growing Rods and a Mode of Distraction Failure

机译:弯曲的杆签署了一种新的放射线标志,以检测磁控生长杆的分心机制中的变形线和分散衰竭的模式

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Study Design. Case report. Objective. To report a unique mechanical failure of magnetically controlled growing rods (MCGRs) that is related to continuous rod slippage. Summary of Background Data. Despite the growing popularity of MCGR in the management of patients with early onset scoliosis, postoperative complications and reoperations are not uncommon. Unique complications or unplanned reoperations are observed in patients with MCGRs that are not seen in traditional growing rods. The complications include rod slippage, mismatch between targeted and achieved distraction length, metallosis, and actuator pin fracture. We have identified an unreported failure mechanism whereby deformed threads occur in the internal distraction mechanism of the MCGR. This phenomenon may indicate increased MCGR internal screw friction and increased distraction resistance. Increased friction and resistance ultimately led to distraction failure and revision surgery. Methods. A girl with early onset scoliosis (EOS) was treated by dual MCGRs. She developed proximal junctional kyphosis (PJK) and continuous rod slippage. Distractions in the out-patient clinic and under sedation did not achieve successful rod lengthening. A tell-tale "crooked rod" radiological sign was identified by angulation between the actuator and the extendable portion of the rod. Results. Rod exchange was performed and she is now 2 years after revision surgery with successful lengthening episodes. Upon review of the extracted MCGR, distraction was not possible even after rod removal and was only resumed after the screw was manually reinserted along its correct threads. Conclusion. Deformed threads is a complication that cannot be reverted without rod removal and exchange. Users should be aware of this potential failure mechanism and proceed to early rod exchange rather than attempting any further distractions.
机译:学习规划。案例报告。客观的。报告与连续杆滑动有关的磁控生长杆(MCGR)的独特机械故障。背景数据摘要。尽管MCGR在早期发病患者的患者中越来越受到普及,但术后并发症和重新进入并不罕见。在传统的生长杆上未见的MCGR患者中观察到独特的并发症或无计划的重新进步。并发症包括杆滑动,目标和达到分心长度,金属异位和致动器销骨折之间的不匹配。我们已经确定了一个未报告的失​​败机制,即在MCGR的内部分心机制中发生变形的线程。这种现象可以表明MCGR内部螺杆摩擦增加并增加了抗干扰性。增加摩擦和抗性最终导致分散注意力和修正手术。方法。有早期发作脊柱侧凸(EOS)的女孩被双MCGRS治疗。她开发了近端通路脊柱脊髓(PJK)和连续杆滑动。在门诊诊所和镇静下的分心并未实现成功的杆延长。通过致动器和杆的可伸展部分之间的角度识别出讲述“弯曲棒”放射学标志。结果。罗德交易所进行了,她现在经过成功延长发作后2年后2年。在审查提取的MCGR时,即使在杆去除后,均不可分取,并且在螺杆沿其正确的螺纹手动重新插入后才恢复。结论。变形螺纹是在没有杆拆卸和交换的情况下无法恢复的并发症。用户应该意识到这种潜在的失败机制,并前进到早期棒交换,而不是尝试进一步的分心。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号