Patient-specific QA and delivery verification of scanned ion beam at NIRS-HIMAC

机译:NIRS-HIMAC的患者特异性QA和扫描离子束的交付验证

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: To evaluate a patient-specific QA program and system for constancy checking of a scanning delivery system developed at the National Institute of Radiological Sciences. Methods: For the patient-specific QA, all the planned beams are recalculated on a water phantom with treatment planning software (TPS). The recalculated dose distributions are compared with the measured distributions using a 2D ionization chamber array at several depths, and evaluated using gamma index analysis with criteria of 3% and 3 mm and a pass rate of 90%. For the constancy check, the authors developed the multiwire proportional chamber (MWPC), which can record the delivered 2D fluence images in a slice-by-slice manner. During irradiation for dosimetric QA with the 2D ionization chamber array and an accordion-type water phantom, the 2D fluence images are recorded using the MWPC in the delivery system. These recorded images are then compared to those taken in the treatment session to check the constancy check. This analysis also employs gamma index analysis using the same criteria as in the patient-specific QA. These patient-specific QA and constancy check evaluations were performed using the data of 122 patients. Results: In the patient-specific QA, the measured dose distributions agreed well with those calculated by the TPS, and the QA criteria were satisfied in all measurements. The additional check of the fluence comparison ensured the constancy of the delivered field during each treatment irradiation. Conclusions: The authors established a patient-specific QA program and additional check of delivery constancy in every treatment session. Fluence comparison is a strong tool for constancy checking of the delivery system.
机译:目的:评估一个特定于患者的QA程序和系统,以便在国家放射科学研究所开发的扫描交付系统的常规检查。方法:对于患者特异性的QA,所有计划的梁在具有治疗计划软件(TPS)的水体系上重新计算。将重新计算的剂量分布与使用2D电离室阵列的测量分布进行比较,并且使用γ指数分析评估,标准为3%和3mm,并且通过90%的通过率。对于恒定检查,作者开发了多励量比例室(MWPC),其可以以切片方式以切片方式记录交付的2D注速图像。在用2D电离室阵列和手风琴式水体模型的剂量测定QA照射期间,使用递送系统中的MWPC记录2D注速器图像。然后将这些记录的图像与处理会话中拍摄的那些进行比较,以检查常规检查。该分析还使用与患者特异性QA中相同的标准使用伽马指数分析。使用122名患者的数据进行这些患者特异性的QA和常规检查评估。结果:在患者特异性QA中,测量的剂量分布与通过TPS计算的那些吻合良好,并且在所有测量中满足QA标准。电量比较的额外检查确保了每种治疗辐照过程中递送的现场的恒定。结论:作者建立了患者特定的QA计划,并在每次治疗会议中额外支票。 Fluence比较是用于跨越输送系统的强大工具。

著录项

  • 来源
    《Medical Physics》 |2013年第12期|共1页
  • 作者单位

    Medical Physics Research Group Research Center for Charged Particle Therapy National Institute of;

    Medical Physics Research Group Research Center for Charged Particle Therapy National Institute of;

    Medical Physics Research Group Research Center for Charged Particle Therapy National Institute of;

    Medical Physics Research Group Research Center for Charged Particle Therapy National Institute of;

    Medical Physics Research Group Research Center for Charged Particle Therapy National Institute of;

    Medical Physics Research Group Research Center for Charged Particle Therapy National Institute of;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类 基础医学;
  • 关键词

    Constancy check; Particle therapy; Patient-specific QA; Scanning delivery;

    机译:常规检查;颗粒疗法;特定患者的QA;扫描交付;

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号