Purpose: To present a clinical case in which real-time intratreatment imaging identified an multileaf collimator (MLC) leaf to be consistently deviating from its programmed and logged position by >1 mm. Methods: An EPID-based exit-fiuence dosimetry system designed to prevent gross delivery errors was used to capture cine during treatment images. The author serendipitously visually identified a suspected MLC leaf displacement that was not otherwise detected. The leaf position as recorded on the EPID images was measured and log-files were analyzed for the treatment in question, the prior day's treatment, and for daily MLC test patterns acquired on those treatment days. Additional standard test patterns were used to quantify the leaf position. Results: Whereas the log-file reported no difference between planned and recorded positions, image-based measurements showed the leaf to be 1.3 ?.1 mm medial from the planned position. This offset was confirmed with the test pattern irradiations. Conclusions: It has been clinically observed that log-file derived leaf positions can differ from their actual position by >1 mm, and therefore cannot be considered to be the actual leaf positions. This cautions the use of log-based methods for MLC or patient quality assurance without independent confirmation of log integrity. Frequent verification of MLC positions through independent means is a necessary precondition to trust log-file records. Intratreatment EPID imaging provides a method to capture departures from MLC planned positions.
机译:目的:介绍一个临床情况,其中实时腹腔成像识别多叶准直器(MLC)叶,以持续偏离其编程和记录位置> 1 mm。方法:旨在防止毛重交付误差的基于杂志的出口 - Fiuence Dosimetry系统用于在治疗图像期间捕获Cine。作者在目前鉴定了未被检测到的疑似的MLC叶片位移。记录在ePID图像上的叶子位置是测量的,分析了对问题的治疗,前一天的治疗以及在那些治疗日中获得的日常生活测试模式的对数文件。其他标准测试模式用于量化叶子位置。结果:虽然日志文件报告计划和录制位置之间没有区别,但基于图像的测量显示叶子为1.3毫米的内侧。通过测试模式照射确认了该偏移量。结论:已经临床观察到,日志文件衍生的叶子位置可以与它们的实际位置不同> 1mm,因此不能被认为是实际的叶子位置。本注意使用基于日志的方法来使用MLC或患者质量保证,而无需独立确认日志完整性。通过独立方式频繁验证MLC位置是信任日志文件记录的必要前提条件。腹部陷入困境成像提供了一种捕获MLC计划位置的偏离的方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号