SU‐F‐J‐78: Characterization of Deformable Image Registration for the Pelvic Region Based On Prostate Shifting, Image Noise and the Existence of Implanted Fiducial Markers

机译:SU-F-J-78:基于前列腺换档,图像噪声和植入基准标记的骨盆区域的可变形图像配准的表征

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Purpose: To assess the impact of prostate shifting, image noise and the existence of fiducial markers on deformable image registration (DIR) performance in the pelvic region using original digital phantoms. Methods: A digital phantom of the pelvic region of a prostate cancer patient was created using contoured anatomy as a base image set. Two simulated fiducial markers were implanted at the apex and base of the prostate. With and without markers, the prostate was shifted by 6.00 and 10.0 mm superior, 5.63 and 9.38 mm posterior, and 5.63 and 9.38 mm left, respectively. We also added image noise with a standard deviation (SD) of 9, 21, or 28. A phantom with a noise SD of 9 and zero prostate shift was defined as the target, and the other phantoms were defined as the sources. DIR was performed using MIM Maestro (MIM Software Inc., Cleveland, OH, USA) on the digital phantoms. A dice similarity coefficient (DSC) was used as the index of correspondence level of the prostates on the target images and the results of the DIR. Also, we calculated the motion distance of the fiducial markers. Results: The DSC decreased with increasing prostate distance. The degree of correspondence of the fiducial markers and the DIR performance were worse when the bladder volume changed after the prostate shifted in the superior direction than when it shifted in the other directions. Changes in the image noise and the existence of fiducial markers produced minimal impact on the DIR results. Conclusion: We found that DIR was not affected by image noise, and the reproduction of prostate position by maintaining the bladder volume was important to preserve DIR performance. Further, the existence of two fiducial markers had no impact on DIR performance. This work was supported by Japan Society for the Promotion of Science (JSPS) KAKEN Grant‐in‐Aid for Scientific Research (C) Grant number 26460724.
机译:目的:评估前列腺转移,图像噪声和基准标记的存在对盆腔区域中可变形图像配准(DIR)性能的影响,使用原始数码模型。方法:使用轮廓解剖学作为基础图像集产生前列腺癌患者的盆腔区域的数模幽灵。将两个模拟的基准标记物植入前列腺的顶点和底座。在没有标记的情况下,前列腺偏移6.00和10.0 mm的后,5.63和9.38毫米,分别为5.63和9.38 mm。我们还添加了具有9,21或28的标准偏差(SD)的图像噪声。具有9个和零前列位移的噪声SD的幻影被定义为目标,并且其他幽灵被定义为来源。使用MIM Maestro(MIM Software Inc.,Cleveland,Oh Usa)在数码幻影上进行DIR。将骰子相似度系数(DSC)用作目标图像上前列腺的对应程度和目录结果的指数。此外,我们计算了基准标记的运动距离。结果:DSC随着前列腺距离的增加而降低。当膀胱体积在前列腺位移在上方方向上的前列位移之后发生时,基准标记和缺血性能的对应程度更差。图像噪声的变化和基准标记的存在对DIR结果产生了最小的影响。结论:我们发现,DIR不影响图像噪声,并保留DIR表现前列腺部位通过保持膀胱容量的再现是很重要的。此外,两个基准标记的存在对Dir表现没有影响。这项工作得到了日本促进科学协会(JSP)的科学研究助手(C)授予第26460724号的助手。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号