Accuracy of a dose-area product compared to an absorbed dose to water at a point in a 2 cm diameter field

机译:与2厘米直径场的点的吸收剂量相比,剂量区域产品的准确性

获取原文
获取原文并翻译 | 示例

摘要

Purpose: Graphite calorimeters with a core diameter larger than the beam can be used to establish dosimetric references in small fields. The dose-area product (DAP) measured can theoretically be linked to an absorbed dose at a point by the determination of a profile correction. This study aims at comparing the DAP-based protocol to the usual absorbed dose at a point protocol in a 2 cm diameter field for which both references exist.
机译:目的:石墨热量计大于光束的芯直径,可用于建立小型字段中的Dosimetric引用。 测量的剂量区域产品(DAP)理论上可以通过确定轮廓矫正的点在一个点上与吸收剂量连接。 该研究旨在将基于DAP的方案与两种直径场中的点协议中的常规吸收剂量进行比较。