To reduce radiation exposure in 320 slice computed tomography (CT), a volume exposure control (VEC) system, is used frequently. The VEC system considers the degree of transmission of radiation, which correlates strongly with the body mass index and the degree of obesity and reduces the tube current. The information is acquired from scout images and its validity is achieved as a reduction of tube current, especially in non-obese subjects. However, if tube current is reduced, image noise may increase, especially in pulsating heart images.
机译:为了减少320片计算机断层扫描(CT)中的辐射曝光,经常使用体积曝光控制(VEC)系统。 VEC系统考虑辐射的传播程度,其强烈地与体重指数和肥胖程度强烈相关,并减少管电流。 从SCOUT图像获取信息,并且其有效性是由于管电流的减少而实现,尤其是在非肥胖的主题中。 然而,如果管电流减小,则图像噪声可能增加,尤其是在脉动心脏图像中。