摘要

Functional carbon nanotubes (CNT) have attracted much attention for analytical and biomedical applications. This paper describes the fabrication of a cholesterol oxidase (ChOx) immobilised polyaniline (PANI)/CNT composite electrode for the amperometric detection of cholesterol. The prepared ChOx/PANI/CNT/Au bioelectrode bound ChOx via the available functionalties of PANI (-NH2) and CNT (-COOH). Moreover, the CNT creates a network inside the matrix that strengthens the mechanical property of the bioelectrode. The multifunctional matrix is presumed to provide a 3D-mesoporous surface, which substantially enhances enzyme activity. The linear range of the biosensor for cholesterol oleate was 30-280 mu M with a response time of 10 sec.
机译:功能性碳纳米管(CNT)已引起分析和生物医学应用的广泛关注。本文介绍了用于胆固醇安培检测的固定胆固醇氧化酶(ChOx)的聚苯胺(PANI)/ CNT复合电极的制备方法。制备的ChOx / PANI / CNT / Au生物电极通过PANI(-NH2)和CNT(-COOH)的可用功能与ChOx结合。此外,CNT在基质内部创建了一个网络,该网络增强了生物电极的机械性能。据推测,多功能基质可提供3D中孔表面,该表面可显着增强酶的活性。胆固醇油酸酯生物传感器的线性范围为30-280μM,响应时间为10秒。