首页> 外文期刊>Journal of chromatography, A: Including electrophoresis and other separation methods >The performance of atmospheric pressure gas chromatography-tandem mass spectrometry compared to gas chromatography-high resolution mass spectrometry for the analysis of polychlorinated dioxins and polychlorinated biphenyls in food and feed samples
【24h】

The performance of atmospheric pressure gas chromatography-tandem mass spectrometry compared to gas chromatography-high resolution mass spectrometry for the analysis of polychlorinated dioxins and polychlorinated biphenyls in food and feed samples

机译:大气压气相色谱-串联质谱与气相色谱-高分辨率质谱相比用于分析食品和饲料样品中多氯二恶英和多氯联苯的性能

获取原文
获取原文并翻译 | 示例
           

摘要

Recently, gas chromatography tandem mass spectrometry (GC-MS/MS) has been added in European Union (EU) legislation as an alternative to magnetic sector high resolution mass spectrometry (HRMS) for the analysis of dioxins and dioxin like polychlorinated biphenyls (dl-PCB) in food and feed. In this study the performance of APGC-MS/MS compared to GC-HRMS is investigated and compared with EU legislation. The study includes the legislative parameters, relative intermediate precision standard deviation (S-RW,rel), trueness, sensitivity, linear range and ion ratio tolerance. In addition, over 200 real samples of large variety and spanning several orders of magnitude in concentration were analyzed by both techniques and the selectivity was evaluated by comparing chromatograms. The S-RW,rel and trueness were evaluated using (in-house) reference samples and fulfill to EU legislation, though the S-RW,rel was better with GC-HRMS. The sensitivity was considerably better than of GC-HRMS while the linear range was similar. Ion ratios were mostly within the tolerable range of +/- 15%. A (temporary unresolved) systematic deviation in ion ratio was observed for several congeners, yet this did not lead to exceeding of the maximum ion ratio limits. The APGC-MS/MS results for the non-dioxin-like-PCBs (ndl-PCBs) were negatively biased, particularly for PCB138 and 153 in contaminated samples. The selectivity of APGC-MS/MS was lower for several matrices. Particularly for contaminated samples, interfering peaks were observed in the APGC chromatograms of the native compounds (dioxins) and labeled internal standards (PCBs). These can lead to biased results and ultimately to false positive samples. It was concluded that the determination of dioxins and PCBs using APGC-MS/MS meets the requirements set by the European Commission. However, due to generally better selectivity and S-RW,rel, GC-HRMS is the preferred method for monitoring purposes. (C) 2016 Elsevier B.V. All rights reserved.
机译:最近,气相色谱串联质谱法(GC-MS / MS)已在欧盟(EU)立法中添加,作为磁性部门高分辨率质谱法(HRMS)的替代方法,用于分析二恶英类和二恶英类多氯联苯(dl-食品和饲料中的多氯联苯)。在这项研究中,对APGC-MS / MS与GC-HRMS的性能进行了比较,并与欧盟法规进行了比较。研究包括立法参数,相对中间精度标准偏差(S-RW,相对),真实性,灵敏度,线性范围和离子比率公差。此外,通过这两种技术分析了200多个种类繁多,浓度跨越几个数量级的真实样品,并通过比较色谱图评估了选择性。 S-RW,rel和真实性是使用(内部)参考样品评估的,并且符合欧盟法规,尽管GC-HRMS的S-RW,rel更好。线性范围相似时,灵敏度明显优于GC-HRMS。离子比大部分在+/- 15%的容许范围内。观察到几个同类物的离子比率(暂时未解决)系统偏差,但这并未导致超过最大离子比率限制。非二恶英样PCBs(ndl-PCBs)的APGC-MS / MS结果有负偏倚,尤其是对污染样品中的PCB138和153。 APGC-MS / MS对几种基质的选择性较低。特别是对于受污染的样品,在天然化合物(二恶英)和标记的内标(PCB)的APGC色谱图中观察到干扰峰。这些会导致结果有偏差,并最终导致假阳性样本。结论是,使用APGC-MS / MS测定二恶英和PCB符合欧洲委员会设定的要求。但是,由于通常具有更好的选择性和S-RWrel,因此GC-HRMS是监测的首选方法。 (C)2016 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号