首页> 外文期刊>Journal of chromatography, A: Including electrophoresis and other separation methods >Fish contamination by polychlorobiphenyls: The mass spectrometric ortho effect in a new and easy gas chromatography-mass spectrometry method for the analysis of the seven indicators. The case of Bluefin Tuna
【24h】

Fish contamination by polychlorobiphenyls: The mass spectrometric ortho effect in a new and easy gas chromatography-mass spectrometry method for the analysis of the seven indicators. The case of Bluefin Tuna

机译:多氯联苯对鱼的污染:一种新的简便气相色谱-质谱法中的质谱邻位效应,用于分析七个指标。蓝鳍金枪鱼案

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

A simple instrumental procedure was developed to carry out the not simple analysis of PCBs in fish samples. PCBs with the same degree of chlorination (the isomers) are expected to be totally indistinguishable among them by all existing detectors and by all existing mass spectrometers, and there is no apparent solution in those frequent cases where two isomers chromatographically coelute. Generally such coelutions are solved by means of multidimensional GC, but it is a complex technique impractical for most laboratories. The present research focuses on the seven important "indicator PCBs" by using the mass spectrometer in an innovative way. The "mass spectrometric ortho effect" was usefully exploited in addressing coelutions between isomers. Other new important observations on mass spectra were decisive in solving the apparent problem arising from coelutions between higher chlorinated PCBs with the lower chlorinated ones when low-resolution MS is used. With the proposed procedure, the seven indicators are analyzed in a simple way and with a degree of accuracy never observed with the conventional gas chromatography. The method was applied to some Bluefin Tuna fish samples of big size suspected to have not negligible levels of PCBs due to the high position of this species in the food chain. The supposition was partly confirmed. On the basis of the results here obtained, the recently introduced EU Regulation on six of the seven indicators shows one critical point: in the present paper, an amendment to the Regulation is proposed. A number of important validation measures are reported. (C) 2014 Elsevier B.V. All rights reserved.
机译:开发了一种简单的仪器程序来进行鱼样品中多氯联苯的简单分析。预期所有现有的检测器和所有现有的质谱仪都无法完全区分具有相同氯化度的PCB(异构体),在两种异构体色谱共洗脱的常见情况下,没有明显的解决方案。通常,此类共洗脱是通过多维气相色谱法解决的,但这对于大多数实验室而言是一项不切实际的复杂技术。本研究通过创新方式使用质谱仪,重点研究了七个重要的“指示器PCB”。在解决异构体之间的共洗脱问题时,可利用“质谱邻位效应”。当使用低分辨率质谱分析时,其他有关质谱的重要新发现对于解决由较高氯化PCB与较低氯化PCB共洗脱引起的明显问题起决定性作用。通过提出的程序,可以以简单的方式分析七个指示剂,并且准确度达到了常规气相色谱法从未观察到的程度。该方法应用于一些大尺寸的蓝鳍金枪鱼鱼样本,由于该物种在食物链中的较高位置,这些样本被怀疑具有不可忽略的多氯联苯含量。该假设得到了部分证实。根据此处获得的结果,最近引入的关于七个指标中六个指标的欧盟法规显示了一个关键点:在本文件中,提出了对该法规的修订。报告了许多重要的验证措施。 (C)2014 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号