首页> 外文期刊>Journal of chromatography, A: Including electrophoresis and other separation methods >Efficiency of short, small-diameter columns for reversed-phase liquid chromatography under practical operating conditions
【24h】

Efficiency of short, small-diameter columns for reversed-phase liquid chromatography under practical operating conditions

机译:在实际操作条件下用于反相液相色谱的小直径短色谱柱的效率

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Prototype small-size (1.0 mm I.D., 5 cm long) columns for reversed-phase HPLC were evaluated in relation to instrument requirements. The performance of three types of columns, monolithic silica and particulate silica (2 mu m, totally porous and 2.6 mu m, core-shell particles) was studied in the presence of considerable or minimal extra-column effects, while the detector contribution to band broadening was minimized by employing a small size UV-detector cell (6- or 90 nL). A micro-LC instrument having small system volume (<1 mu L) provided extra-column band variance of only 0.01-0.02 mu L-2. The three columns generated about 8500 theoretical plates for solutes with retention factor, k> 1-3 (depending on the column), in acetonitrile/water mobile phase (65/35 = vol/vol) at 0.05 mL/min, with the instrument specified above. The column efficiency was lower by up to 30% than that observed with a 2.1 mm 1.12 commercial column. The small-size columns also provided 8000-8500 theoretical plates for well retained solutes with a commercial ultrahigh-pressure liquid chromatography (UHPLC) instrument when extra-column contributions were minimized. While a significant extra-column effect was observed for early eluting solutes (k<2-4, depending on column) with methanol/water (20/80 = vol/vol) as weak-wash solvent, the use of methanol/water = 50/50 as wash solvent affected the column efficiency for most analytes. The results suggest that the band compression effect by the weak-wash solvent associated with partial-loop injection may provide a practical means to reducing the extra-column effect for small-size columns, while the use of an instrument with minimum extra-column effect is highly desirable. (C) 2015 Elsevier B.V. All rights reserved.
机译:根据仪器要求评估了用于反相HPLC的原型小尺寸(内径1.0 mm,长5 cm)。研究了三种类型的色谱柱的性能:整体硅胶和颗粒硅胶(2微米,完全多孔,2.6微米,核-壳型颗粒),在存在明显或最小的外柱效应的情况下,而检测器对谱带的贡献通过使用小尺寸的紫外线检测器电池(6-或90 nL)可以将加宽最小化。系统体积较小(<1μL)的微型液相色谱仪提供的柱外谱带差异仅为0.01-0.02μL-2。这三根色谱柱在乙腈/水流动相(65/35 = vol / vol)中以0.05 mL / min的流速生成了约8500个理论板的保留因子为k> 1-3(取决于色谱柱)的溶质以上指定。与2.1 mm 1.12商业色谱柱相比,色谱柱效率降低了多达30%。小型色谱柱还提供了8000-8500的理论塔板,可在商业上使用超高压液相色谱(UHPLC)仪器对保留好的溶质进行分析,从而最大程度地减少了柱外的影响。对于使用甲醇/水(20/80 =体积/体积)作为弱洗溶剂的早期洗脱溶质(k <2-4,取决于色谱柱),观察到了显着的柱外效应,而使用甲醇/水= 50/50的洗涤溶剂会影响大多数分析物的色谱柱效率。结果表明,与部分回路进样相关的弱洗溶剂的带压缩效应可能为减少小尺寸色谱柱的柱外效应提供了一种实用的方法,同时使用了具有最小柱外效应的仪器是非常可取的。 (C)2015 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号