Cyanides are poisonous chemicals that widely exist in nature and industrial processes as well as accidental fires. Rapid and accurate determination of cyanide exposure would facilitate forensic investigation, medical diagnosis, and chronic cyanide monitoring. Here, a rapid and direct method was developed for the determination of cyanide ions in urinary samples. This technique was based on an integrated capillary electrophoresis system coupled with laser-induced fluorescence (LIF) detection. Cyanide ions were derivatized with naphthalene-2,3-dicarboxaldehyde (NDA) and a primary amine (glycine) for LIF detection. Three separate reagents, NDA, glycine, and cyanide sample, were mixed online, which secured uniform conditions between samples for cyanide derivatization and reduced the risk of precipitation formation of mixtures. Conditions were optimized; the derivatization was completed in 2-4min, and the separation was observed in 25s. The limit of detection (LOD) was 4.0 nM at 3-fold signal-to-noise ratio for standard cyanide in buffer. The cyanide levels in urine samples from smokers and non-smokers were determined by using the method of standard addition, which demonstrated significant difference of cyanide levels in urinary samples from the two groups of people. The developed method was rapid and accurate, and is anticipated to be applicable to cyanide detection in waste water with appropriate modification. Published by Elsevier B.V.
机译:氰化物是有毒化学物质,广泛存在于自然和工业过程以及意外火灾中。快速准确地确定氰化物的暴露量将有助于法医调查,医学诊断和长期氰化物监测。在此,开发了一种快速直接的方法来测定尿样中的氰化物离子。该技术基于集成的毛细管电泳系统以及激光诱导荧光(LIF)检测。用萘-2,3-二甲醛(NDA)和伯胺(甘氨酸)衍生氰离子进行LIF检测。在线混合三种独立的试剂NDA,甘氨酸和氰化物样品,这确保了氰化物衍生化样品之间的均匀条件,并降低了混合物沉淀形成的风险。条件已优化;衍生化过程在2-4分钟内完成,分离时间为25秒。缓冲液中标准氰化物的信噪比为3倍时,检出限(LOD)为4.0 nM。采用标准添加法测定了吸烟者和非吸烟者尿液样本中的氰化物含量,这表明两组人尿液样本中的氰化物含量存在显着差异。所开发的方法快速而准确,并有望通过适当的修改将其应用于废水中氰化物的检测。由Elsevier B.V.发布

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号