首页> 外文期刊>Journal of chromatography, A: Including electrophoresis and other separation methods >Determination of phenylalanine in blood by high-performance anion-exchange chromatography-pulsed amperometric detection to diagnose phenylketonuria
【24h】

Determination of phenylalanine in blood by high-performance anion-exchange chromatography-pulsed amperometric detection to diagnose phenylketonuria

机译:高效阴离子交换色谱-脉冲安培检测法测定血液中的苯丙氨酸,以诊断苯丙酮尿症

获取原文
获取原文并翻译 | 示例
           

摘要

We have developed a high-performance anion-exchange chromatography with pulsed amperometric detection method for the detection of phenylalanine (Phe) and diagnosis of phenylketonuria (PKU). Sample pretreatment steps were simplified without derivatization. The analyte was separated within 5 min. The detection limit (S/N = 3) for Phe was 50 pg. Linear dynamic range was 1.23-14.43 mg/dL (r(2) = 0.9999) for a dried blood spot. The mean recoveries of Phe for intra- and inter-clay assays were found to be 96.87-104.16%. This method clearly differentiated PKU-positive groups from normal groups, and proved to be a practical procedure for rapid screening and follow-up monitoring of PKU.
机译:我们开发了一种具有脉冲安培检测方法的高效阴离子交换色谱法,用于检测苯丙氨酸(Phe)和诊断苯丙酮尿症(PKU)。简化了样品预处理步骤而没有衍生化。在5分钟内分离出分析物。 Phe的检出限(S / N = 3)为50 pg。干血斑的线性动态范围为1.23-14.43 mg / dL(r(2)= 0.9999)。进行粘土内和粘土间测定的Phe平均回收率为96.87-104.16%。该方法清楚地将PKU阳性组与正常组区分开,并被证明是用于PKU快速筛查和随访监测的实用程序。