Purpose Some patients fitted with an AMS 800? type artificial urinary sphincter report urinary leaks in the sitting position whose mechanism remains unclear. In this experimental study we explain these residual leaks by characterizing fluid transfer and pressure variations in the device during cuff compression. Materials and Methods Two complete sets of AMS 800 devices were tested on an original test bench. A qualitative study using methylene blue dye was performed to assess fluid transfer within the device during manual cuff compression. After estimating the pressures exerted on the perineal floor in the sitting position, pressure variations in the AMS 800 components were measured in real time using the pressure sensors of a urodynamic workstation. Results When the AMS 800 cuff was compressed between 2 fingers, reflux of the blue dye into the balloon through the pump resistor was observed. After the cuff was compressed for 5, 10, 15 and 20 seconds at a pressure of 150 cm H2O, the minimal pressure measurements in the cuff were 37, 25, 14 and 12 cm H 2O, respectively. Furthermore, the time required for cuff re-pressurization was 68, 87, 101 and 117 seconds, respectively. Conclusions Urinary leaks reported in sitting position by patients fitted with an AMS 800 may be explained by compression of the urethral cuff between the pelvic floor and the chair, which causes emptying of the cuff into the balloon. Patients should be reassured that there is no implant dysfunction and may adopt simple lifestyle strategies to circumvent these leaks.
机译:目的一些装有AMS 800的患者?人工尿道括约肌类型报告坐位尿漏,其机制尚不清楚。在本实验研究中,我们通过表征袖带压迫期间设备中的液体传输和压力变化来解释这些残留泄漏。材料和方法在原始测试台上测试了两套完整的AMS 800设备。进行了使用亚甲蓝染料的定性研究,以评估手动袖带压迫期间器械内的液体转移。估计在坐姿时会阴地板上施加的压力后,使用尿动力学工作站的压力传感器实时测量AMS 800组件中的压力变化。结果当将AMS 800套囊压在两个手指之间时,观察到蓝色染料通过泵电阻器回流到气球中。在将袖带在150 cm H2O的压力下压缩5、10、15和20秒后,袖带中的最小压力测量值分别为37、25、14和12 cm H 2O。此外,袖带再加压所需的时间分别为68、87、101和117秒。结论装有AMS 800的患者坐姿报告的尿液渗漏可能是由于骨盆底和椅子之间的尿道套囊受压引起的,这导致套囊排空到气球中。应确保患者没有植入物功能障碍,并可以采取简单的生活方式策略来规避这些渗漏。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号